<p>Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park Twinsburg, OH 44087-2371 Phone: 330.425.1313 / 800.800.3106 Fax: 330.405.8064 www.hitachimed.com AIRIS II Standard Details Revised: March 8, 2005 It is the customer's responsibility to ensure that all appropriate design and construction personnel receive this document along with their site specific drawing. This document contains information which is critical to the proper design and construction of a suite for the Hitachi AIRIS MRI system. The information contained herein is to be used in conjunction with the site specific drawing provided by HMSA. Contact the Site Planning Department with questions regarding this guide. Contents: Sheet 1 of 17: Typical Planning Process Sheet 10 of 17: Lighting/Environmental Notes Sheet 2 of 17: Typical Project Schedule Sheet 11 of 17: Gauss Fields Sheet 3 of 17: Floor Plans Sheet 12 of 17: Magnet Placement Sheet 4 of 17: General Siting Considerations Sheet 13 of 17: Rigging Procedure Sheet 5 of 17: RF Shielding Information Sheet 14 of 17: Rigging Details Sheet 6 of 17: Structural Notes Sheet 15 of 17: Components - Scan Room Sheet 7 of 17: Electrical - General Sheet 16 of 17: Components - Control Room Sheet 8 of 17: Electrical-Grounding Notes Sheet 17 of 17: Preliminary & Final Checklists Sheet 9 of 17: System Wireway Information OEM Addendum DOCUMENT ESTABLISHED - 1/03/00 SPD-GDE-AIRIS</p><p>AIRIS II REVISIONS MARCH 8, 2005 (PREVIOUS GUIDE: JANUARY 3, 2000) THIS REVISED GUIDE REFLECTS MINOR MODIFICATIONS OF THE AIRIS STANDARD DETAILS. PLEASE NOTE THE FOLLOWING SPECIFIC MODIFICATIONS THAT SHOULD BE REVIEWED BY YOUR DESIGN PERSONNEL. THE CUSTOMER IS RESPONSIBLE FOR PREPARING THEIR AIRIS SUITE ACCORDING TO THE REQUIREMENTS PRESENTED HERE. PG. 3 OF 17: FLOOR PLANS 1) CHANGED "EQUIPMENT/ACCESSORIES BY OTHERS" LABELING FROM LETTERS TO NUMBERS PG. 5 OF 17: RF SHIELDING INFORMATION 1) RF SHIELDING NOTES - 6.G.: RF ROOM VERIFICATION TEST REQUIRED AFTER SYSTEM DELIVERY - UPDATED NOTE AT END OF RF SHIELD NOTES PG. 9 OF 17: SYSTEM WIREWAY INFORMATION 1) ADDED NOTE 2 TO WIREWAY NOTES 2) REVISED FILTER PANEL (SIDE VIEW) TO REFLECT AIR GAP PG. 10 OF 17: LIGHTING/ENVIRONMENTAL NOTES 1) REVISED NOTE 1 (SCAN ROOM TEMPERATURE) OF ENVIRONMENTAL NOTES 2) REVISED NOTE 7 OF ENVIRONMENTAL NOTES 3) REVISED ENVIRONMENTAL DIAGRAM TO REFLECT DIELECTRIC BREAKS 4) ADDED NO TRACK LIGHTING NOTE PG. 15 OF 17: COMPONENTS SCAN ROOM 1) REVISED FILTER PANEL (SIDE VIEW) TO REFLECT AIR GAP PG. 16 OF 17: COMPONENTS CONTROL ROOM 1) ADDED NETWORK INFORMATION 2) ADDED POWER CONDITIONING INFORMATION PG. 17 OF 17: AIRIS II CHECKLISTS 1) REVISED SCAN ROOM TEMPERATURE IN AIRIS SITE CHECKLIST FOR INSTALLATION OEM 1) ADDED SAFETY FERROGUARD TO VENDOR LIST 2) UPDATED AIRIS II PAINT CODES HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 (800) 800-3106 REVISIONS DATE: 01/03/00 REVISION: 03/08/05 AIRIS STANDARD DETAILS</p><p>TYPICAL PLANNING PROCESS SITE SELECTION AND EVALUATION (4 WEEKS) E) INFORMATION CONSOLIDATION AND EVALUATION (1 WEEK) D) PRE-CONSTRUCTION MEETING (SCHEDULED BY CUSTOMER/ARCHITECT) A) SITE SELECTION (1 WEEK) 1) SITE PLANNING WILL: ALL PLANS REVIEWED AND OUTSTANDING QUESTIONS ARE A CONSTRUCTION TIME 1) THE SALES PERSON WILL: REVIEW ALL CHECKLIST AND DRAWING INFORMATION LINE SHOULD BE AGREED UPON AT THIS MEETING AND A DELIVERY TIME THE FIELD SUBMIT PARTIALLY COMPLETED SITE SELECTION SURVEY WITH CUSTOMER REVIEW ALL TEST RESULTS WITH TECH SUPPORT SERVICE ENGINEER WILL ALSO BE INTRODUCED TO THE CUSTOMER AND INFORMATION CONSULT WITH SALES AND FIELD SERVICE IF REQUIRED CONTRACTOR AT THIS THE ENGINEER WILL BEGIN TIMELY SITE EVALUATION VISITS - REQUEST EVALUATION ASSISTANCE FROM FIELD SERVICE GENERATE SITE EVALUATION REPORT AND DISCUSS ACTIONS AND RECOMMENDATIONS BEGINNING WITH THIS 2) SITE PLANNING WILL: WITH SALES RECORD DATE OF REQUEST AND FORWARD REQUEST OF ASSISTANCE TO DISTRICT MANAGER SITE CONSTRUCTION (6-8 WEEKS) FORWARD SITE PLANNING GUIDE TO CUSTOMER IF REQUESTED CUSTOMER REVIEW AND PRELIMINARY PLANS (VARIES) A) ON SITE EVALUATION 3) DISTRICT MANAGER WILL: A) SITE EVALUATION REPORT (1 DAY) THE FIELD SERVICE ENGINEER WILL REVIEW WORKS IN PROGRESS, VERIFYING SITE SCHEDULE FIELD SERVICE VISIT WITH SALESPERSON FOR SITE EVALUATION SITE PLANNING WILL FORWARD SITE EVALUATION REPORT TO SALES FOR PRESENTATION TO CONSTRUCTION PROGRESS AND ADVISING SITE PLANNING OF BOTH SITE STATUS AND ANY B) SITE EVALUATION CUSTOMER CONCERNS SITE PLANNING WILL ADDRESS CONCERNS WITH ARCHITECT OR CONTRACTOR AS 1) THE SALES PERSON WILL: B) REVIEW AND RETURN OF SITE EVALUATION REPORT SUBMIT A COMPLETED SITE SELECTION SURVEY SALES WILL REVIEW THE REPORT WITH THE CUSTOMER (SITE PLANNING IS AVAILABLE FOR B) CONSTRUCTION PRIOR TO DELIVERY PROVIDE A SCALED DRAWING OR DIMENSIONED SKETCH OF THE PROPOSED SITE TELEPHONE CONSULTATION) AND RETURN A SIGNED COPY WITH THE CUSTOMERS ACCEPTANCE THE FOLLOWING MUST BE IN PLACE PRIOR TO DELIVERY OF 2) SITE PLANNING WILL: C) SITE DRAWING AND DETAILS (2-4 WEEKS) DEMOLITION 4 WEEKS RECORD DATE OF REVIEW INFORMATION AND IDENTIFY SITING CONCERNS C) PRELIMINARY DRAWING AND TEST REQUEST (1 WEEK) SITE PLANNING WILL PREPARE A DETAILED LAYOUT AND PROCESS THE SITE CONCERNS, FRAMING OF THE SITE 3 WEEKS THIS MAY INCLUDE: RF ROOM 1 WEEK (NOTE LEAD TIME WHEN ORDERING RF SHIELD) 1) SITE PLANNING WILL: MAGNETIC SHIELD DESIGN FOR FRINGE FIELD CONTROL NOTE: THE RF ROOM WILL VARY BETWEEN VENDORS, THIS COULD ADD TIME TO - PREPARE A PRELIMINARY SKETCH WITH A CLEARLY IDENTIFIED MAGNET ISOCENTER MAGNETIC SHIELD DESIGN FOR CONTROL OF INTERFERENCE - E-MAIL A SKETCH, A COMPLETED AIRIS PRELIMINARY SITE SELECTION SURVEY, AND A D) ON SITE MEETING (SCHEDULED BY CUSTOMER) LEAVE ACCESS FOR DELIVERY OF SYSTEM FIELD SERVICE ENGINEER SITE SURVEY TO THE DISTRICT MANAGER (2) 120 VAC, 20 AMP SINGLE PHASE OUTLETS IN THE SCAN TESTED 2) DISTRICT MANAGER WILL: SALES AND SITE PLANNING TO MEET WITH THE CUSTOMER AND CUSTOMER'S ARCHITECT. DELIVERY STAGING AREA MUST BE IDENTIFIED FORWARD INFORMATION TO THE APPROPRIATE FIELD SERVICE ENGINEER THE INTENTION OF THIS MEETING IS TO: CONTROL ROOM LIGHTING AT LEAST TEMPORARY NOTE: TESTING IS INVALID WITHOUT A CLEARLY MARKED VERIFY PREVIOUS CHECKLIST INFORMATION LIGHTING MUST BE AVAILABLE IN THE SCAN ROOM (BY CONTRACTOR). D) SITE TESTING (1 WEEK) REVIEW AND FINALIZE THE EQUIPMENT LAYOUT DELIVERY SCHEDULE STORAGE AREA (IF REQUIRED) CLEAR REVIEW STANDARD DETAILS 1) FIELD SERVICE WILL: RIGGING GEAR AND FIELD ENGINEER SCHEDULED ADDRESS OUTSTANDING CUSTOMER QUESTIONS OR CONCERNS SCHEDULE AND COMPLETE SITE TEST FOR MAGNETIC (AC AND DC) INTERFERENCE C) PRE-DELIVERY CONFERENCE (DICTATED BY CONSTRUCTION SCHEDULE) IF THERE IS AN EXISTING MRI IN THE AREA, PLOT MAGNETIC FIELD LINES FIELD ENGINEER WORKS WITH SITE RIGGER, RF VENDOR, AND CONTRACTOR TO VERIFY PREVIOUSLY SUBMITTED SITE INFORMATION PRECONSTRUCTION COORDINATE MAGNET VERIFY MINIMUM DELIVERY STANDARDS ARE MET AND A SKETCH SITE OR MODIFY SITE PLANNING E-MAILED SKETCH OF SITE, NOTING ALL POTENTIAL SOURCES OF INTERFERENCE A) ARCHITECT PREPARES CONSTRUCTION DRAWINGS (6-8 WEEKS) UNOBSTRUCTED PATH IS COMPLETE FIELD SERVICE ENGINEER SURVEY INPUT MUST BE RECEIVED FROM: D) DELIVERY (1 DAY) HMSA DELIVERY AND POSITIONING OF MAGNET AT SITE E-MAIL INFORMATION TO MR TESTING GROUP FOR PROCESSING ENGINEERS (STRUCTURAL, HVAC, ELECTRICAL) INVENTORY AND STORAGE OF ANCILLARY EQUIPMENT AT SITE NOTE: 1) ALL TESTING MUST BE DONE DURING PERIODS OF TYPICAL ACTIVITY IN THE RF SHIELDING VENDOR NOTE: IF A SAFE, SECURABLE STORAGE AREA IS NOT AVAILABLE AT YOUR SITE, PLEASE NOTE SURROUNDING AREA. RIGGER THE FOLLOWING: THE HMSA FIELD ENGINEER AND THE RIGGER WILL UNLOAD COMPONENTS NEEDED FOR 2) ACTIVITY IN THE IMMEDIATE AREA OF THE ISOCENTER MUST BE RESTRICTED B) HMSA REVIEW OF ARCHITECTURAL DRAWINGS (1 WEEK) SITE DURING TESTING. SITE PLANNING WILL DO A CURSORY REVIEW OF THE ARCHITECTURAL NOTE NORTH AMERICAN VAN LINES WILL STORE THE ANCILLARY COMPONENTS OFF SITE. 3) HMSA COMPENSATION UNIT CAN CORRECT FOR ONLY A SINGLE SOURCE OF AND ADVISE THE ARCHITECT THIS IS A REVIEW AND NOT AN APPROVAL HMSA A REQUIRED SECOND DELIVERY WILL BE SCHEDULED UPON SUITE AC AND A SINGLE SOURCE OF DC MULTIPLE SOURCES WOULD ACCEPTS NO LIABILITY FOR ERRORS OR IT is THE RESPONSIBILITY OF THE CUSTOMER ALL COSTS ASSOCIATED WITH A SECOND DELIVERY ARE THE RESPONSIBILITY OF THE REQUIRE: AND THEIR DESIGN AND CONSTRUCTION PROFESSIONALS TO PREPARE THEIR SITE TO MEET THE THE SITE TO BE RELOCATED GUIDELINES CONTAINED IN THIS DOCUMENT AND IN ACCORDANCE WITH THE SITE TO HAVE SOME OF THE SOURCES ELIMINATED C) GENERAL CONSTRUCTION (4 WEEKS) THE CUSTOMER TO USE MAGNETIC SHIELDING HITACHI MEDICAL SYSTEMS, AMERICA THE DRAWINGS GO OUT TO BID AND A GENERAL CONTRACTOR WILL BE CHOSEN BY THE 1959 SUMMIT COMMERCE PARK OH 44087-2371 4) THERE ARE LIMITATIONS ON THE AMOUNT OF INTERFERENCE THAT THE CUSTOMER THE CUSTOMER/ARCHITECT MUST NOTIFY SITE PLANNING WITH A CONTACT AND (800) 800-3106 COMPENSATION UNIT CAN CORRECT FOR. TELEPHONE THE GENERAL CONTRACTOR WILL ALSO BE RESPONSIBLE FOR SECURING TYPICAL PLANNING PROCESS THE BUILDING PERMITS DURING THIS DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 1 OF 17</p><p>TYPICAL PROJECT SCHEDULE SITE CONSTRUCTION - CONTINUED E) PRE-INSTALLATION CONSTRUCTION (2-3 WEEKS) SEE "SITE CHECKLIST FOR INSTALLATION" (pg. 17 of 17) ALL ROOM FINISHES (CEILING, WALLS, FLOORING) COMPLETED ALL ELECTRICAL REQUIREMENTS (PANEL, LIGHTS, OUTLETS) COMPLETED J-BOX FOR MRI CABINET IN PLACE W/10' PIGTAIL WIREWAY INSTALLED HVAC SYSTEM OPERATIONAL AND TESTED RF ROOM COMPLETED (ACCESS OPENING SECURED, FILTER PANEL INSTALLED) AND TESTED TYPICAL PROJECT SCHEDULE SUITE IS CLEAN, FREE OF DUST AND DEBRIS NOTE: DURING FINISH THE MAGNET WILL BE AT FIELD, THEREFORE NON-FERROUS ACTIVITY MONTH1 MONTH2 MONTH3 MONTH4 MONTH5 MONTH6 TOOLS ARE W1 W1 W1 SITE SELECTION AND EVALUATION BY SALES INSTALLATION (3 WEEKS) PRELIMINARY DRAWING BY SITE PLANNING INSTALLATION WILL NOT COMMENCE UNTIL COMPLETED CHECKLIST IS RECEIVED BY HMSA SITE TESTING BY HMSA NO CONSTRUCTION MAY TAKE PLACE IN THE SCAN AND CONTROL ROOM DURING TEST EVALUATION BY HMSA INSTALLATION SALES REVIEW OF EVALUATION WITH CUSTOMER CONTRACTOR AND ELECTRICIAN ALL MUST BE AVAILABLE DURING INSTALLATION HMSA PREPARES DETAIL SITE SPECIFIC DRAWINGS DICOM II, MODEM, AND CAMERA OUTLETS MUST BE INSTALLED AND OPERATIONAL ON SITE MEETING W/ / KEY CUSTOMER PERSONNEL CUSTOMER'S ARCHITECT PREPARES CONSTRUCTION DRAWINGS APPLICATIONS (1 WEEK) HMSA REVIEW OF CONSTRUCTION DOCUMENTS OBTAIN AND AWARD CONSTRUCTION BID AN HMSA APPLICATIONS SPECIALIST WILL WORK WITH YOUR TECHNOLOGISTS TO FAMILIARIZE THEM SUBMISSION OF DRAWINGS TO CITY FOR PERMITS WITH THE VARIED OPERATIONS AND PROCEDURES OF THE AIRIS II AN ENTIRE WEEK OF INITIAL APPLICATIONS SHOULD BE SET ASIDE FOR TRAINING, WHICH INCLUDES A LIMITED SCHEDULE PRIMARY SITE CONSTRUCTION WORK OF SCANNING VOLUNTEERS AND AIRIS DELIVERY PRE-INSTALLATION CONSTRUCTION COMPLETION INSTALLATION NOTE: APPLICATIONS 1) SITE PLANNING IS AVAILABLE FOR TELEPHONE CONSULTATION AT ALL TIMES THROUGHOUT THIS PROCESS. 2) STANDARD DETAILS WILL BE PROVIDED AT ANY TIME. 3) ALL INDIVIDUALS INVOLVED IN THE CONSTRUCTION OF YOUR SUITE SHOULD BE PROVIDED WITH THIS PLANNING GUIDE AS WELL AS A COPY OF THE FINAL HMSA DRAWING. 4) THE ABOVE ESTIMATED TIME FRAME SHOULD BE AMPLE TIME TO COMPLETE THE SITE AND ENSURE A TROUBLE FREE DELIVERY AND INSTALLATION. THIS TIME FRAME WILL VARY FROM SITE TO SITE AND IS AN ESTIMATE ONLY. HMSA WILL NOT BE HELD ACCOUNTABLE FOR ANY CONSTRUCTION DELAYS DUE TO CONTRACTORS NOT SUPPLIED BY HMSA. PLEASE NOTE THAT SOME STEPS CAN OCCUR CONCURRENTLY AS INDICATED ON THE ADJACENT GANT CHART. HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK OH 44087-2371 (800) 800-3106 WWW.HITACHIMED.COM TYPICAL PROJECT SCHEDULE DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 2 OF 17</p><p>HITACHI AIRIS II DRAWING LEGEND RF SHIELD ARCHITECT TO VERIFY WALL CONSTRUCTION WITH RF VENDOR. IT MUST PROVIDE 80db ATTENUATION OR BETTER FROM 0.5-30MHz. (TYPICAL 5" RF WALL THICKNESS REFLECTED.) VIEW WINDOW, TYPICALLY 4'w X 3'h, MOUNTED ABOVE FINISHED FLOOR REMOVABLE EQUIPMENT ACCESS PANEL FOR DELIVERY 16'-0" OF THE SYSTEM. TYPICALLY 8'w X 7'h. 7'-0" 15'-0" EQUAL EQUAL EQUAL EQUAL JUNCTION BOX FOR MRI POWER 5 5 120V, 20 AMP CONVENIENCE AT LEAST ONE REQUIRED FOR SERVICE AND MAINTENANCE H H D 4 6'-6" 3 3 TWO (2) PHONE LINES REQUIRED IN CONTROL ROOM. ONE MIN E \MIN WITH DIRECT OUTSIDE ACCESS FOR REMOTE DIAGNOSTICS AT MRI CABINET OTHER FOR USE BY OPERATOR A AND SERVICE 4 A DATA PORT FOR IN-HOUSE NETWORK CONNECTION WITH MINIMUM OF CATEGORY 5 CABLE. CUSTOMER TO IDENTIFY CONTROL ROOM AND CONFIRM REQUIRED CONNECTION (8'-0" TYPICAL HT.) 19'-4" 20'-0" F F 16" ON-WALL WIREWAY WITH REMOVABLE COVER AND DIVIDERS. METAL ONLY. MUST BE SCAN ROOM IN SCAN ROOM. (8'-0" B 5 GAUSS B 5 GAUSS SCAN ROOM 16" ON-FLOOR WIREWAY WITH REMOVABLE COVER 3G AND METAL ONLY. 3G 1G 1G G MAGNET ISOCENTER G EQUIPMENT BY HMSA E 20' MECHANICAL ELECTRICAL EXCLUSION ZONE A AIRIS GANTRY F SHIM PLATES MECHANICAL ELECTRICAL EXCLUSION ZONE B PATIENT TABLE G COIL STORAGE CABINET 4 OPERATOR'S WORKSTATION H WALL MOUNTED SPEAKER 6 D MRI CABINET 9'-6" CONTROL ROOM E FILTER PANEL ALTERNATIVE FLOOR PLAN - 425 SQ. FT. (8'-0" TYPICAL CLG HT.) "NOT TO SCALE" EQUIPMENT/ACCESSORIES BY OTHERS 1 OPTIONAL UPS UNIT 2 OPTIONAL LASER IMAGER TYPICAL FLOOR PLAN - 480 SQ. FT. 3 4" X 16" NON-FERROMAGNETIC ON-FLOOR WIREWAY "NOT TO SCALE" 4 4" X 16" NON-FERROMAGNETIC ON-WALL WIREWAY 5 DELIVERY ACCESS PANEL HITACHI MEDICAL SYSTEMS, AMERICA 6 CHAIR 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 7 CASEWORK (800) 800-3106 FLOOR PLANS NOTE: NOT ALL COMPONENTS ARE REFLECTED ON PLANS. DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 3 OF 17</p><p>GENERAL SITING CONSIDERATIONS AIRIS II EQUIPMENT LEGEND SITE SELECTION FOR PROPER EVALUATION OF A POTENTIAL SITE, A SET OF DRAWINGS OF THE PROPOSED MRI LOCATION AND THE SURROUNDING COMPONENT QTY WIDTH DEPTH HEIGHT WEIGHT (lbs) BTUH AREA SHOULD BE SENT TO HMSA SITE PLANNING FOR REVIEW. THE FOLLOWING ITEMS SHOULD BE DISCUSSED WITH THE SITE A. GANTRY 1 9'-0" 5'-11" 6'-5" PLANNING SPECIALIST FOR A THOROUGH EVALUATION OF A PROPOSED LOCATION (ALSO SEE pg. 11 of 17): 34,615 1706 1) LOCATION OF ANY MEDICAL IMAGING EQUIPMENT (CT, X-RAY, MAMMOGRAPHY, ETC.) IMMEDIATELY ADJACENT TO, OR B. PATIENT TABLE 1 2'-9" 7'-9" 3'-0" 882 120 BELOW THE PROPOSED MRI SCAN ROOM. C. OPERATOR'S WORKSTATION 1 4'-0" 2'-6" 2'-6" 73 800 2) LOCATION OF FRINGE FIELDS (PARTICULARLY THE 1 GAUSS FIELD) OF ANY EXISTING OR FUTURE MRI. 1. MONITOR 1 1'-8" 1'-10" 1'-7" 17 N/A 3) LOCATION OF EXISTING OR FUTURE AIR CONDITIONING UNITS (AIR HANDLERS OR CONDENSER) WITHIN 20 FEET OF PROPOSED 2. MOD UNIT 1 0'-5" 0'-10" 1'-0" 18 N/A MAGNET 4) LOCATION OF ANY STRUCTURAL STEEL BEAMS AND/OR COLUMNS WITHIN 10 FEET IN ALL DIRECTIONS OF THE PROPOSED 3. KEYBOARD 1 1'-8" 0'-10" 0'-3" 5 N/A MAGNET D. MRI CABINET 1 3'-4" 2'-7" 5'-10" 1213 6230 5) LOCATION OF ANY VEHICULAR TRAFFIC (INCLUDING PARKING) WITHIN 20 FEET OF THE PROPOSED MAGNET E. FILTER PANEL 1 2'-3" 6"-20" 3'-9" 133 N/A 6) LOCATION OF ELECTRICAL SERVICE (50 AMPS OR HIGHER), ANY ELECTRICAL OR MECHANICAL ROOMS AND ELEVATORS WITHIN F. SHIM PLATES 2 4'-7" 4'-7" 1/8" 40 EA. N/A 40 FEET OF THE PROPOSED MAGNET 2 4'-7" 4'-1" 1/8" 40 EA. N/A 7) LOCATION OF ANY LARGE MOVING FERROUS OBJECTS (TRAINS, SUBWAYS, ETC.) IN THE IMMEDIATE G. COIL STORAGE CABINET 1 5'-0" 2'-3" 4'-4" 50 N/A NOTE: COORDINATION BETWEEN THE ARCHITECT, CONTRACTORS, AND THE HMSA SITE PLANNING SPECIALIST SHOULD BEGIN VERY H. SPEAKER 1 0'-6" 0'-6" 0'-8" 3 N/A EARLY IN THE PLANNING PROCESS. OPTIONAL COMPONENTS BY CUSTOMER - EQUIPMENT WILL VARY PLANNING 1. *UPS UNIT 1 1'-7" 2'-3" 12" 440 2500 TO MAXIMIZE THE PERFORMANCE OF THE AIRIS II MRI SYSTEM, CAREFUL PLANNING SHOULD TAKE PLACE FROM THE INITIAL SITE 2. *LASER IMAGER 1 SEE HMSA SALES FOR PRODUCT OPTIONS SELECTION STAGE, THROUGH DESIGN AND CONSTRUCTION, AND THROUGHOUT THE FINAL DELIVERY AND INSTALLATION PROCESSES. MAGNETIC COMP. UNIT 1 16-1/2" 13-1/2" 1'-2" 49 10 PROPER PLANNING WILL HELP TO ENSURE HIGH SYSTEM PERFORMANCE STANDARDS, COMFORTABLE PATIENT HANDLING, AN EFFICIENT MAGNETIC COMP. COIL 1 6'-6" 6'-6" 3-1/2" 55 N/A WORKING AND SUFFICIENT STORAGE SPACE. IN GENERAL, THE FOLLOWING SHOULD BE CONSIDERED WHEN DESIGNING AN MRI SUITE: (*) SEE HMSA SALES FOR DETAILS 1) THE 5 GAUSS FIELD SHOULD BE CONTAINED WITHIN THE MRI SCAN ROOM. 2) THE MRI SCAN ROOM SHOULD BE LOCATED NEAR THE EXTERIOR OF THE BUILDING WHEN POSSIBLE TO FACILITATE DELIVERY PATIENT COMFORT AND HANDLING (SEE RIGGING INFORMATION pg. 13 of 17). TO FACILITATE PATIENT HANDLING AND PROVIDE FOR PATIENT SAFETY, THE FOLLOWING SHOULD BE CONSIDERED IN THE 3) CONVENIENCE OUTLETS SHOULD BE LOCATED IN THE MRI SCAN ROOM FOR SERVICE. DESIGN OF THE MRI SUITE: 4) THE CONTROL ROOM SHOULD BE SUFFICIENTLY LARGE TO ACCOMMODATE EMPLOYEE ACTIVITIES AND TO PROVIDE FOR 1) PATIENT ACCESS INTO THE MRI SCAN ROOM SHOULD ACCOMMODATE A STRETCHER. SERVICE ACCESS TO ALL THE 2) THERE SHOULD BE SUFFICIENT AREA AROUND THE PATIENT TABLE IN THE SCAN ROOM FOR PATIENT 5) THE UPS UNIT (IF PURCHASED) AND THE LASER IMAGER CAN BE LOCATED REMOTELY. SEE LASER IMAGER AND DICOM 3) DRESSING ROOMS, TOILETS, HOLDING AREAS, PREP ROOMS, SHOULD BE LOCATED NEAR THE MRI SCAN ROOM. PLANNING CONCERNS (pg. 16 of 17). 4) THE OPERATOR SHOULD HAVE DIRECT ACCESS TO THE OPERATIONS FLOOR SPACE IN DESIGNING AN MRI SUITE THAT IS FUNCTIONAL AND THE FOLLOWING SHOULD BE CONSIDERED: THE AIRIS SYSTEM HAS BEEN SPECIFICALLY DESIGNED TO BE SITED WITHIN AN EXISTING FACILITY. THE STANDARD SYSTEM 1) STRONG VISUAL CONTACT BETWEEN THE OPERATOR AND THE PATIENT SHOULD BE CONSISTS OF ONLY SIX (6) MAIN COMPONENTS, WHICH CAN BE COMFORTABLY SITED WITHIN 425 SQUARE FEET. 2) SUFFICIENT AND CONVENIENT WORK SURFACES AND STORAGE SPACE SHOULD BE INCLUDED IN AND NEAR THE MRI SCAN AND CONTROL ROOMS. MR MAGNETIC FIELD 3) A SINK SHOULD BE LOCATED NEARBY (BUT NOT IN THE MRI SCAN ROOM) FOR THE OPERATOR. AS WITH ALL MR SYSTEMS, A MAJOR CONCERN IN PLANNING IS ACCOMMODATING THE SYSTEM'S MAGNETIC FIELD. THIS FIELD 4) A TELEPHONE WITH ACCESS TO AN OUTSIDE LINE SHOULD BE LOCATED IN THE CONTROL ROOM TO FACILITATE APPLICATION WILL AFFECT MAGNETICALLY SENSITIVE INSTRUMENTS AND BE AFFECTED BY FERROMAGNETIC MATERIALS IN THE SURROUNDING SUPPORT FOR THE OPERATOR AND TECHNICAL SUPPORT FOR SERVICE AREA. THE SIZE OF THE FRINGE FIELD IS DETERMINED BY THE STRENGTH AND TYPE OF THE MAGNET. SINCE THE AIRIS SYSTEM 5) AN OFFICE INTERCOM SHOULD BE LOCATED IN THE CONTROL ROOM FOR THE SAFETY OF THE PATIENT AND THE OPERATOR. INCORPORATES A SELF-SHIELDED PERMANENT MAGNET, ITS ASSOCIATED FRINGE FIELD IS MINIMAL (pg. 11 of 17). IN TYPICAL 6) WHILE THE LASER IMAGER AND PROCESSOR CAN BE REMOTE, IT SHOULD BE LOCATED TO BEST MEET YOUR OPERATIONAL SITING, THE 5 GAUSS FRINGE FIELD WILL BE CONTAINED WITHIN THE SCAN ROOM, ELIMINATING ADDITIONAL MAGNETIC SHIELDING NEEDS (pg. 16 of 17). COSTS. 7) THE AIRIS HAS A DIGITAL OUTPUT. YOUR LASER IMAGER SHOULD BE SELECTED HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK OH 44087-2371 (800) 800-3106 WWW.HITACHIMED.COM GENERAL SITING CONSIDERATIONS DATE: 01/03/00 03/08/05 AIRIS II STANDARD DETAILS SHT 4 OF 17</p><p>16'-0" RF SHIELD NOTES EQUAL EQUAL 2 1) AN RF SHIELDED ROOM IS REQUIRED FOR PROPER OPERATION OF YOUR MRI SYSTEM. YOUR HMSA SITE PLANNER CAN PROVIDE YOU WITH THE NAMES OF A NUMBER OF QUALIFIED RF VENDORS (SEE OEM ADDENDUM FOR LIST OF VENDORS). 2) THE RF SHIELDED ROOM MUST HAVE AN INSULATION RESISTANCE OF MORE THAN 1,000 OHMS (AS MEASURED WITH A 4'-7 500 V-DC, MEGGER) WITH RESPECT TO THE GROUNDING LINE. 3) THE RF SHIELDED ROOM MUST HAVE AN INSULATION RESISTANCE OF MORE THAN 1,000 OHMS (AS MEASURED WITH 6 6 A 500 V-DC MEGGER) WITH RESPECT TO THE BUILDING GROUNDING FACILITY AND OTHER EQUIPMENT INSTALLED IN THE 6'-6" BUILDING. MIN. 4) THE RF SHIELD MUST PROVIDE 80db ATTENUATION OR BETTER FROM 0.5 TO 30MHz. 4'-7 5) CONSIDERATION MUST BE GIVEN TO THE ISOLATION OF THE RF SHIELD WITH RESPECT TO THE FIRE SPRINKLER THE PIPES MUST BE ELECTRICALLY ISOLATED FROM THE RF SHIELD AND A "DRY" OR PREACTION FIRE SPRINKLER SYSTEM MAGNET IS A SINK SHOULD NOT BE LOCATED IN THE MRI SCAN ROOM. FEET 6) THE RF VENDOR IS RESPONSIBLE FOR, AND YOUR QUOTE SHOULD REFLECT THE FOLLOWING: A) INSTALLATION OF THE HMSA FILTER PANEL. FILTER PANEL MOUNTING METHODS WILL 4'-1 DIFFER. CONTACT RF VENDOR FOR ROUGH OPENING REQUIREMENTS (SEE pg. 15 of 17). ISOCENTER B) INSTALLATION OF THE HMSA SHIM PLATES. THE PLATES ARE INCORPORATED WITHIN ALL VENDORS' 6 THE VENDOR MUST INSTALL THE SHIM PLATES DURING THE INSTALLATION OF THEIR FLOOR. C) SUPPLY AND INSTALL ELECTRICAL FILTERS (EMI) FOR LIGHTS AND OUTLETS IN THE MRI SCAN ROOM D) SUPPLY AND INSTALL RF DOOR (4'-0"w X 7'-0"h TYPICAL) 1 E) SUPPLY AND INSTALL RF WINDOW (4'-0"w X 3'-0"h TYPICAL) SCAN F) SUPPLY AND INSTALL WAVEGUIDES FOR HVAC SUPPLY AND RETURN DUCTS IN THE MRI SCAN ROOM ROOM G) PERFORM AN RF ROOM VERIFICATION TEST AFTER SYSTEM DELIVERY AND PROVIDE THE CUSTOMER WITH A WRITTEN CONFIRMATION TO BE MADE AVAILABLE TO SITE PLANNING UPON REQUEST H) DETAILED RF SHIELD INSTALLATION DRAWINGS 7) THE RF SHIELDING VENDOR MUST BE CONTACTED IF A SOUND SYSTEM IS TO BE INSTALLED IN THE MRI SCAN ROOM. A FILTER FOR THE SPEAKERS MUST BE PROVIDED. THE SPEAKERS MUST BE LOCATED OUTSIDE THE 1 GAUSS FIELD. 5 3 8) WHEN POSSIBLE, THE CONCRETE SLAB IN THE MRI SCAN ROOM SHOULD BE RECESSED TO ACCOMMODATE THE THICKNESS 4 OF THE RF IF THE SLAB IS NOT RECESSED, A SMALL RAMP IS REQUIRED AT THE THRESHOLD TO THE SCAN ROOM. 9) METHODS OF CONSTRUCTION AND MATERIALS USED WILL VARY AMONG RF MANUFACTURERS. THEREFORE AN RF VENDOR MUST BE SELECTED PRIOR TO CONSTRUCTION TO INSURE THAT THE MRI SCAN ROOM AREA IS PREPARED PROPERLY. NOTE: PLACEMENT OF WAVEGUIDES (FOR HVAC) AND EMI FILTERS MUST BE COORDINATED BETWEEN THE RF VENDOR AIRIS II RF SHIELDING LEGEND AND ARCHITECT/CONTRACTOR TO BEST MEET YOUR SITE NEEDS. WHILE THE WAVEGUIDES ARE PROVIDED BY THE RF VENDOR, THE CONNECTION OF THE DUCTWORK, INCLUSIVE OF THE DIELECTRIC BREAK, IS THE RESPONSIBILITY OF THE HVAC CONTRACTOR. 1 RF SHIELD FOUR (4) WALLS, AND FLOOR REQUIRED TO PROVIDE 80db ATTENUATION FROM 0.5 TO 30MHz. TYPICAL 5" WALL THICKNESS SHOWN WILL VARY WITH VENDOR. 2 ACCESS PANEL: 8'-0"w X 7'-0"h CLEAR OPENING REQUIRED FOR DELIVERY OF THE MAGNET. LOCATION MAY VARY. 3 FILTER PANEL FOR SYSTEM PROVIDED BY HMSA AND INSTALLED BY RF VENDOR PRIOR TO FINAL ROOM TEST. 4 RF WINDOW: 4'-0"w X 3'-0"h (TYPICAL) MTD. AT 36" A/FF 5 RF DOOR: 4'-0"w X 7'-0"h (TYPICAL) HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 6 FOUR (4) SHIM PLATES, PROVIDED BY HMSA, INSTALLED BY RF (800) 800-3106 VENDOR (SEE pg. 12 of 17 FOR SHIM PLATE DETAILS) RF SHIELDING INFORMATION DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 5 OF 17</p><p>9'-0" STRUCTURAL AND FLOOR LOADING INFORMATION G 1'-6 1'-11 1'-11 15, COMPONENT 1'-6 WEIGHT (lbs) LOADING PATTERN 16 MAGNET 34,615 3 POINT DISTRIBUTION: 13,000 lbs FRONT FOOT 10,800 lbs REAR FOOT, EACH PATIENT TABLE 882 (2) FLOOR PLATES AS SHOWN REAR REAR TABLE LEGS FOOT FOOT OPER. WORKSTATION 73 SEE PG. 16 OF 17 3'-5" (4) 1.75" 2.5" DIA. WHEELS AND (4) 2.5" DIA. LEVELING JACKS MRI CABINET 1213 SEE PG. 16 OF 17 ISOCENTER 5'-11" FILTER PANEL WALL MOUNTED BY RF VENDOR 133 G SEE PG. 15 OF 17 8 16 STRUCTURAL NOTES 2'-3 1'-6 FRONT FOOT 1) ALL STRUCTURAL SUPPORT IS THE CUSTOMER'S A REVIEW OF BOTH THE DELIVERY PATH AND THE MRI SCAN 1" ROOM LOCATION MUST BE MADE BY A QUALIFIED PROFESSIONAL TO ENSURE ADEQUATE SUPPORT IS WHEN DESIGNING A SUPPORT CARE MUST BE TAKEN TO SUPPORT THE MAGNET "JACKING LOCATIONS" (SHOWN ON pg. 14 of 17). EACH JACK MAY NEED TO SUPPORT ONE-HALF THE MAGNET'S WEIGHT DURING DELIVERY. 1'-8" 1'-6 15, 16 13'-9" 2) THE HMSA SHIM PLATES ARE DESIGNED FOR MAGNETIC FIELD HOMOGENEITY ONLY, AND ARE NOT INTENDED FOR STRUCTURAL 3) THE CONTRACTOR IS RESPONSIBLE TO PROVIDE A FLAT AND LEVEL SLAB FOR THE RF VENDOR'S SHIELDED FLOOR. THE SLAB 3'-10 3. SHOULD BE LEVEL TO 1/4" OVER 10 FEET AND FLAT TO 1/10" PER SQUARE FOOT. IN NEW THE SLAB IN THE MRI 10'-4" 2'-2 SCAN ROOM SHOULD BE RECESSED TO ACCOMMODATE THE THICKNESS OF THE RF SHIELDED FLOOR AND THE HMSA SHIM PLATES. TABLE BASE 4) A VINYL TILE FLOOR IS RECOMMENDED THROUGHOUT THE MRI SUITE FOR EASE OF SERVICE AND OPERATION. CARPET IN THE PLATES SCAN AND CONTROL ROOMS IS DISCOURAGED DUE TO STATIC BUILD-UP, DUST DIFFICULTY MOVING EQUIPMENT 7'-9" FOR SERVICING, AND SANITARY THE FLOOR COVERING THICKNESS SHOULD NOT EXCEED YOUR HMSA SITE PLANNING SPECIALIST AND RF VENDOR MUST BE NOTIFIED PRIOR TO RF SHIELD INSTALLATION IF THE FLOOR THICKNESS 1'-4 EXCEEDS 5) THE MAGNET ROOM SLAB SHOULD BE ISOLATED FROM THE SURROUNDING AREA USING UNISORB OR LIKE MATERIAL (MIN. THICKNESS). CONTACT SITE PLANNING REGARDING NON-SLAB ON GRADE CONDITIONS. (SEE OEM ADDENDUM PAGE 3 13, 16 16 FOR VENDOR CONTACT.) TABLE ANCHORS LOCATION NOTE: THE MAGNET ONCE WILL ACT AS THE KEY FOR YOUR PLANNING. THIS REFLECTS THE LOCATION TABLE ANCHORS LOCATION 2" 2" USED DURING INITIAL SITE QUALIFICATION (TESTING), ALL SHIELDING IS REFERENCED TO THIS POINT, THE RF VENDOR POSITIONS THE SHIM PLATES ACCORDINGLY, AND LASTLY, THE RIGGER WILL REFERENCE THIS LOCATION WHEN SETTING YOUR SYSTEM. YOUR STRUCTURE SHOULD BE DESIGNED WITH THE ISOCENTER AS A REFERENCE POINT. 2'-5" 2'-9" HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK OH 44087-2371 (800) 800-3106 PLAN VIEW: FLOOR LOADING "NOT TO SCALE" STRUCTURAL NOTES DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 6 OF 17</p><p>ELECTRICAL - GENERAL NOTES ALL CONSTRUCTION SHALL BE PERFORMED IN COMPLIANCE WITH ALL APPLICABLE LOCAL AND NATIONAL ELECTRIC CODES (NEC) AND FEED FOR LIGHTS AND OUTLETS 1) POWER REQUIREMENTS: 200, 208, 220, 240 VAC, (+5% UNDER LOAD CONDITIONS) NOT TO IN THE MRI SCAN ROOM. CONTACT RF SHIELDING VENDOR FOR DETAILS. EXCEED 240V; 60Hz (+1%); SINGLE-PHASE; 50 AMP (208V, 60 AMP SERVICE IS REQUIRED ONLY NON-CONDUCTIVE CONDUIT OR DIELECTRIC CONNECTORS IF A UPS IS PURCHASED BY THE CUSTOMER). 200/208/220/240V (+/- 50A; FILTERS EMI FILTERS BY RF VENDOR SINGLE-PHASE FOR OUTLETS AND LIGHTS 2) A UPS IS STRONGLY RECOMMENDED FOR ALL SITES IN AREAS WHERE THERE IS A HIGH 3-WIRE WITH INSULATED GROUND FROM MAIN SERVICE ENTRANCE IN MRI SCAN ROOM TO BUILDING MAIN INCIDENT OF LIGHTNING STORMS, POOR QUALITY POWER, OR FREQUENT BLACK/BROWN OUTS. OR ISOLATION TRANSFORMER SERVICE OR ISOLATION 3) THE BUILDING GROUND CONDUCTOR MUST BE 100 OHMS OR LESS TO EARTH TRANSFORMER OUTPUT TO MRI CABINET: 208VAC MRI PROVIDED BY HMSA SUB-PANEL 1) HMSA IS RESPONSIBLE FOR ALL AC POWER WIRING FROM THE MRI CABINET TO THE MRI SYSTEM ALL OTHER WIRING AND MATERIALS WILL BE THE RESPONSIBILITY INDIVIDUAL BRANCH CIRCUITS FOR CONTROL ROOM OUTLETS AND LIGHTS GROUND TO HMSA FILTER PANEL OF THE ELECTRICAL CONTRACTOR. (BY ELECTRICAL CONTRACTOR) LOCAL DISCONNECT 2) HMSA IS RESPONSIBLE FOR ALL GROUND CONDUCTORS FROM THE HMSA FILTER (LOCATED IN CONTROL ROOM) PANEL TO THE MRI SYSTEM INPUT: 208V; 60 AMPS ONLY IF UPS IS PURCHASED RF SHIELDED OPTIONAL SHUNT TRIP MRI SCAN ROOM (BY ELECTRICAL CONTRACTOR) 3) HMSA WILL MAKE ALL SYSTEM CABLE RUNS IN THE 4" X 16" WIREWAY PROVIDED BY THE ELECTRICAL CONTRACTOR. THIS BECOMES THE CUSTOMER'S RESPONSIBILITY SHOULD GROUNDING STUD (BY RF VENDOR) UNION LABOR OR OTHER FACTORS PROHIBIT INSTALLATION BY HMSA PERSONNEL. HMSA MRI CABINET PROVIDED BY THE ELECTRICAL CONTRACTOR (IN MRI CONTROL ROOM) RF WINDOW THE ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR SUPPLYING AND INSTALLING THE FOLLOWING: OPTIONAL UPS HMSA 1) CIRCUIT BREAKER PANEL (MRI SUB-PANEL) IN THE MRI CONTROL ROOM. THIS PANEL SHALL (REMOTE LOCATION) FILTER MRI SCAN PANEL ROOM SERVE ONLY THE FOLLOWING LOADS: MRI CABINET, MRI SUITE RECEPTACLES AND LIGHTS. JUNCTION BOX PIGTAIL IN FLEXIBLE NO OTHER EQUIPMENT INCLUDING THE LASER IMAGER/PROCESSOR SHALL BE CONNECTED (BY ELECTRICAL CONTRACTOR) NON-CONDUCTIVE CONDUIT TO THIS PANEL. THE MRI SUB-PANEL SHALL BE A 120/208 VOLT OR 120/240 VOLT, SINGLE (BY ELECTRICAL CONTRACTOR) -GROUND CONDUCTOR MRI CONTROL (BY HMSA) RF DOOR PHASE WITH 100 AMP MAIN LUGS, BOLT-ON TYPE BRANCH CIRCUIT BREAKERS, AND ROOM ISOLATED GROUND BUS. 10) INSTALLATION OF OPTIONAL UPS SYSTEM IF PURCHASED. 2) PANEL FEEDER SHALL BE COPPER CONDUCTORS SIZED AS REQUIRED BY NATIONAL ELECTRICAL 11) BRANCH CIRCUIT WIRING FOR LASER IMAGER/PROCESSOR PER MANUFACTURER'S CODE (NEC) AND AS REQUIRED TO PROVIDE REQUIRED VOLTAGE LEVEL TO THE MRI FEEDER SPECIFICATIONS FROM BUILDING DISTRIBUTION SYSTEM AS REQUIRED PER NEC SHALL BE FROM A CIRCUIT BREAKER OR FUSED SWITCH SIZED PER NEC IN THE MAIN DISTRIBUTION 12) 4" X 16" WIREWAY WITH REMOVABLE COVER IN MRI SCAN AND CONTROL ROOMS FOR MRI ELECTRICAL ISOMETRIC PANEL AT THE SERVICE ENTRANCE OR THE SOURCE OF THE SEPARATELY DERIVED SYSTEM. "NOT TO SCALE" SYSTEM CABLES. WIREWAY IN SCAN ROOM MUST BE 3) BRANCH CIRCUIT WIRING FROM MRI SUB-PANEL TO MRI CABINET, RECEPTACLES (SEE pg. 9 of 17 FOR REQUIREMENTS.) AND LIGHTS IN MRI SCAN SUITE THROUGH RF FILTERS AND RECEPTACLES IN MRI CONTROL ROOM. 13) 2" 2" WIREWAY WITH REMOVABLE COVER IN CONTROL ROOM FOR SIGNAL CABLE FROM EACH CIRCUIT SHALL BE PROVIDED WITH INDEPENDENT NEUTRAL AND ISOLATED GROUNDING OPERATOR'S WORKSTATION TO LASER IMAGER/PROCESSOR (NETWORK MAY BE USED). CONDUCTORS NOT SHARED WITH OTHER 14) 1/2" WIREMOLD IN SCAN ROOM FOR HMSA WALL MOUNTED SPEAKER (SEE pg. 9 of 17). 4) SAFETY DISCONNECT (WITH OPTIONAL SHUNT TRIP) A LOCAL DISCONNECT WITHIN THE MRI CABINET IS NOT EASILY ACCESSED. A SAFETY DISCONNECT SHOULD BE PROVIDED PROVIDED BY RF SHIELDING VENDOR ADJACENT TO THE MRI CABINET WITH 10 FOOT PIGTAIL IN FLEXIBLE NON-CONDUCTIVE CONDUIT 1) THE VENDOR IS RESPONSIBLE FOR SUPPLYING AND INSTALLING ELECTRICAL FOR CONNECTION TO MRI CABINET. (EMI) FILTERS AT EACH LOCATION AN ELECTRICAL FEED PASSES THROUGH THE RF 5) LIGHTING AND CONTROLS IN MRI SCAN ROOM AND CONTROL ROOM (SEE pg. 10 of 17 FOR SHIELDING. THE LOCATION AND NUMBER OF THESE FILTERS MUST BE COORDINATED WITH TYPICAL LIGHTING PLAN AND NOTES). THE ELECTRICAL CONTRACTOR. 6) 120 VOLT RECEPTACLES WITH ISOLATED GROUND TERMINALS IN MRI SCAN AND CONTROL 2) THE VENDOR IS RESPONSIBLE FOR SUPPLYING AND INSTALLING RF WAVEGUIDES AT EACH ROOMS (SEE pg. 9 of 17 FOR TYPICAL REQUIREMENTS). LOCATION WHERE AN HVAC DUCT PASSES THROUGH THE RF SHIELDING. LOCATION AND HITACHI MEDICAL SYSTEMS, AMERICA 7) ISOLATED GROUNDING SYSTEM FOR ALL LOADS FED BY MRI SUB-PANEL (SEE pg. 8 of 17). 1959 SUMMIT COMMERCE PARK OH 44087-2371 NUMBER OF THESE WAVEGUIDES MUST BE COORDINATED WITH THE HVAC CONTRACTOR 8) RF SHIELD GROUNDING WITH #4 AWG COPPER GROUNDING CONDUCTOR FROM RF SHIELD (800) 800-3106 WWW.HITACHIMED.COM (SEE pg. 10 of 17 FOR RESTRICTIONS). GROUND STUD TO THE HMSA FILTER PANEL. ELECTRICAL GENERAL 3) THE VENDOR IS RESPONSIBLE FOR SUPPLYING AND INSTALLING A GROUND STUD ON 9) CONDUIT AND ENCLOSURE GROUNDING FOR ALL FEEDER AND BRANCH WIRING PER NEC THE RF SHIELD. THIS GROUND STUD SHALL BE THE COMMON GROUND FOR THE MRI DATE: 01/03/00 REVISION: 03/08/05 REQUIREMENTS (SEE GROUNDING NOTES pg. 8 of 17). THE RF SHIELD, AND ALL SCAN ROOM SERVICE. AIRIS STANDARD DETAILS SHT 7 OF 17</p><p>MRI SUB-PANEL MAIN SERVICE EQUIPMENT (SERVICE ENTRANCE) OR TO BE LOCATED IN SEPARATELY DERIVED SOURCE (ISOLATION TRANSFORMER) MRI CONTROL ROOM ISOLATED GROUND ISOLATED GROUND BUS NEUTRAL HOT HOT TYPICAL ELECTRICAL SCHEMATIC HOT 20/1 NEUTRAL INDIVIDUAL BRANCH CIRCUIT TO CONTROL ROOM NEUTRAL BUS ISOLATED GROUND ROOM OUTLETS GROUND STUD G EQUIPMENT AND LIGHTS (BY RF CONTRACTOR) GROUND BUS DIELECTRIC CONNECTOR HOT 4 AWG GROUND 20/1 NEUTRAL INDIVIDUAL BRANCH CIRCUIT EMI FILTER (BY CONTRACTOR) G NEC REQUIRED NEUTRAL-GROUND ISOLATED GROUND TO SCAN ROOM OUTLETS (BY RF VENDOR) HMSA FILTER e PANEL BOND RACEWAY & EQUIPMENT (ONE PLACE ONLY) GROUNDING DIELECTRIC CONNECTOR o HOT TO WORKSTATION NEUTRAL LOCATION AND QUANTITY OF EMI (BY HMSA) 20/1 INDIVIDUAL BRANCH CIRCUIT EMI FILTER G FILTERS ARE TO BE DETERMINED ISOLATED GROUND TO SCAN ROOM LIGHTS (BY RF VENDOR) BY THE RF VENDOR AND CONTRACTOR o TO PATIENT TABLE (BY HMSA) DIELECTRIC CONNECTOR FILTER PANEL o TO GANTRY GROUND BUS- (BY HMSA) HOT o Q 20/1 NEUTRAL G INDIVIDUAL BRANCH CIRCUIT EMI FILTER ISOLATED GROUND TO SCAN ROOM LIGHTS (BY RF VENDOR) G DIELECTRIC TO MRI CAB MAIN BUILDING GROUND CONNECTOR (BY HMSA) HOT HOT HOT o o 6 HOT HOT HOT INDIVIDUAL BRANCH o o 50/2 * NEUTRAL NEUTRAL NEUTRAL CIRCUIT FOR MRI CABINET o ISOLATED GROUND ISOLATED GROUND ISOLATED GROUND COLD WATER PIPE UPS (IF PURCHASED) SERVICE DISCONNECT OR SHUNT TRIP JUNCTION BOX MRI CABINET NEUTRAL BUS ISOLATED FROM PANEL FRAME REQUIRED IF UPS IS PURCHASED WITH 10' PIGTAIL (HMSA PREFERRED) BUILDING STEEL EQUIPMENT GROUND BUS ISOLATED FROM PANEL FRAME (TO BE LOCATED IN CONTROL ROOM) GROUNDING ELECTRODE (<100 ohm IMPEDANCE) 60/2 IF UPS IS PURCHASED (HMSA PREFERRED) GROUNDING NOTES ISOLATED GROUNDING SYSTEM SHALL BE PROVIDED FOR ALL LOADS CONNECTED TO MRI SUB-PANEL AS FOLLOWS: C) METALLIC WIREWAY SYSTEM SHALL BE ISOLATED FROM MRI CABINET AND ELECTRICAL (EMI) FILTERS WITH 1) AN ISOLATED GROUND BUS SHALL BE PROVIDED IN THE MRI SUB-PANEL. THIS BUS SHALL BE INSULATED FROM THE DIELECTRIC CONNECTORS OR CONDUIT. EQUIPMENT GROUNDING SHALL BE ISOLATED FROM PANEL METALLIC WIREWAY WIREWAY TO BE GROUNDED TO FILTER PANEL VIA GROUND STRAP. 2) AN INSULATED COPPER CONDUCTOR SHALL BE RUN FROM THE SYSTEM MAIN GROUND BUS (THE POINT WHERE THE 5) RF SHIELD GROUNDING SHALL BE PROVIDED BY INSTALLATION OF #4 AWG COPPER CONDUCTOR RUN FROM RF ROOM NEUTRAL CONDUCTOR OF THE ELECTRICAL SERVICE OR SEPARATELY DERIVED POWER SOURCE IS BONDED TO THE GROUND STUD TO GROUND BUS IN HMSA FILTER PANEL. GROUNDING OF BUS IN FILTER PANEL TO MRI CABINET WILL BE GROUNDING ELECTRODE CONDUCTOR) TO THE ISOLATED GROUND BUS IN THE MRI SUB-PANEL. THIS CONDUCTOR PROVIDED BY HMSA. SHALL BE THE SAME SIZE AS NEUTRAL CONDUCTOR AND SHALL BE RUN IN CONDUIT ALONG WITH THE FEEDER 6) ELECTRICAL SYSTEM GROUNDING SHALL BE VERIFIED BY THE ELECTRICAL THE SYSTEM NEUTRAL CONDUCTORS SUPPLYING THE PANEL. SHALL BE GROUNDED AT ONE POINT ONLY AT THE SERVICE OR AT THE SOURCE OF THE SEPARATELY 3) AN INDIVIDUAL INSULATED COPPER CONDUCTOR SHALL BE RUN FROM THE ISOLATED GROUND BUS FOR EACH CIRCUIT DERIVED SYSTEM BY CONNECTION TO A NEC REQUIRED GROUNDING ELECTRODE SYSTEM. ORIGINATING IN THE MRI SUB-PANEL TO THE GROUND TERMINAL OF THE LOAD SERVED. GROUNDING CONDUCTORS 7) GROUNDING ELECTRODE SYSTEM SHALL BE VERIFIED AND SUPPLEMENTED IF NECESSARY BY THE ELECTRICAL SHALL BE THE SAME SIZE AS NEUTRAL CONDUCTORS AND RUN IN CONDUIT WITH BRANCH CIRCUIT THE GROUNDING ELECTRODE SYSTEM SHALL BE PROVIDED AS REQUIRED BY THE NEC AND SHALL 4) RACEWAY AND ENCLOSURE GROUNDING SHALL BE PROVIDED AS REQUIRED BY NATIONAL ELECTRICAL CODE (NEC) INCLUDE THE FOLLOWING WHERE AVAILABLE BONDED TOGETHER: METAL UNDERGROUND WATER SERVICE PIPE, FOR ALL FEEDER AND BRANCH CIRCUIT WIRING AS FOLLOWS: METAL FRAME OF BUILDING AND OTHER ELECTRODES REQUIRED BY NEC. IF NOT A) ALL FEEDER AND BRANCH CIRCUIT CONDUCTORS SHALL BE INSTALLED IN STEEL CONDUIT AND BOXES TO SHIELD THE ELECTRICAL CONTRACTOR SHALL PROVIDE ONE OR MORE DRIVEN GROUND RODS OR HITACHI MEDICAL SYSTEMS, AMERICA CONDUCTORS FROM MAGNETIC PLATES ELECTRODES BONDED TO THE EXISTING GROUNDING ELECTRODE SYSTEM BY 1959 SUMMIT COMMERCE PARK OH 44087-2371 (800) 800-3106 B) METALLIC CONDUIT AND BOXES SHALL BE BONDED TOGETHER BY MEANS OF APPROVED CONNECTORS INSTALLED MEANS OF A CONDUCTOR SIZED PER NEC. THE RESISTANCE TO GROUND OF THE MADE IN ACCORDANCE WITH NEC REQUIREMENTS WITH BONDING JUMPERS WHERE NECESSARY TO PROVIDE A ELECTRODE SHALL BE NOT MORE THAN 100 OHMS. ELECTRICAL GROUNDING NOTES CONDUCTIVE GROUNDING PATH FROM EACH OUTLET BACK TO MAIN SYSTEM GROUND AT ELECTRICAL SERVICE DATE: 01/03/00 REVISION: 03/08/05 OR SOURCE OF SEPARATELY DERIVED SYSTEM. AIRIS STANDARD DETAILS SHT 8 OF 17</p><p>WIREWAY NOTES 1) METALLIC WIREWAYS (RACEWAYS) ARE REQUIRED FOR THE SYSTEM WALL MOUNTED SPEAKER BY HMSA CABLES BETWEEN THE VARIOUS MRI WIREWAYS ARE NOT PROVIDED WITH THE SYSTEM AND ARE THE RESPONSIBILITY OF THE A 4" X 16" WALL MOUNTED WIREWAY IS SHOWN 4" X 16" 2'-8" HERE IN A TYPICAL LAYOUT. THE WIREWAY IN THE MRI SCAN ROOM NON-FERROMAGNETIC ON-FLOOR WIREWAY TYPICAL MUST BE NON-FERROMAGNETIC (I.E. ALUMINUM OR STAINLESS STEEL). 4" X 16" NON-FERROMAGNETIC ALL WIREWAYS SHOULD BE MOUNTED AFTER THE FILTER PANEL IS ON-FLOOR WIREWAY INSTALLED. THE WIREWAYS SHOULD BE ORDERED WELL IN ADVANCE WALL MOUNTED SPEAKER RF DOOR -1/2" DUE TO WOOD AND PVC WIREWAYS ARE NOT WIREMOLD BY OTHERS RF WINDOW 2) THE WIREWAY DIVIDERS MUST BE CAPABLE OF SUPPORTING FOUR PHONE MODEM (4) POUNDS PER LINEAR FOOT. FILTER 3) THE WIREWAY MUST HAVE A REMOVABLE COVER AND INCLUDE AT LEAST PANEL SCAN ROOM TWO DIVIDERS CREATING THREE OR MORE 4) AN ON-FLOOR OR IN-FLOOR WIREWAY BEHIND THE GANTRY IS OUTLETS- IT MUST BE REMOVABLE OR PROVIDE AT LEAST 10" CLEARANCE FROM 4" X 16" NON-FERROMAGNETIC THE BACK OF THE GANTRY SO THAT THE GANTRY COVER CAN BE 4" X 16" ON-WALL WIREWAY ON-WALL WIREWAY SCAN ROOM OPEN TRANSITION AREA 5) ALTERNATIVE METHODS OF CABLING MAY BE USED (IN-FLOOR WIREWAY, COMPUTER FLOOR). CABLE LENGTH RESTRICTIONS MAY REQUIRE THAT CONTROL ROOM ALTERNATE METHODS BE USED. THESE MUST BE REVIEWED WITH YOUR SITE PLANNING SPECIALIST PRIOR TO CONSTRUCTION. TYPICAL WIREWAY ISOMETRIC 6) ALL WIREWAYS MUST BE INSTALLED BEFORE HMSA CAN BEGIN THE "NOT TO SCALE" INSTALLATION PROCESS OF THE MRI SYSTEM. 7) A 1/2" NON-FERROMAGNETIC WIREMOLD MUST BE PROVIDED BY THE 4' 16" FINISHED WALLS NON-FERROMAGNETIC ELECTRICAL CONTRACTOR FOR THE HMSA WALL MOUNTED SPEAKER. THE ON-WALL WIREWAY SPEAKER IS TYPICALLY WALL MOUNTED DIRECTLY BEHIND THE GANTRY, 27" NON-CONDUCTIVE FRAMING ABOVE THE 4' X 16" 8) ELECTRICAL CONTRACTOR SHALL PROVIDE A WIREWAY FROM THE RE 23-5/8" SEAL FILTER OPERATOR'S WORKSTATION TO THE LASER IMAGER IN ACCORDANCE ROUGH RF SEAL BOTH SIDES PANEL 35-7/16" WITH THE LASER IMAGER MANUFACTURER'S SPECIFICATIONS (SEE OEM OPENING (1 1/2" AVAILABLE) RF SEAL BOTH SIDES ROUGH 44" (1 1/2" AVAILABLE) ADDENDUM FOR VENDOR CONTACT). OPENING FILTER PANEL COVER RF 4" SEAL 4" X 16" ON-WALL WIREWAY 4" RF WALL 4" 4" 28" 28" CONTROL ROOM 4" X 16" WALL MOUNTED WIREWAY RF FLOOR RF FLOOR 16" 4" FLUSH WITH BOTTOM OF FILTER PANEL TO ROUGH OPN'G TO ROUGH OPN'G WIREWAY 8" 4" SCAN ROOM 4' CONTROL ROOM 4" RF FLOOR- AIR GAP TYPICAL WALL CROSS-SECTION THROUGH FILTER PANEL FOUR 4" X 4" COMPARTMENTS 3 COMPARTMENTS FILTER PANEL FRONT VIEW FILTER PANEL SIDE VIEW TYPICAL FLOOR PLAN (STACKED) (AS NOTED) (VERTICAL ORIENTATION) (VERTICAL ORIENTATION) "NOT TO SCALE" WALL MOUNTED WIREWAY FILTER PANEL TO WIREWAY TRANSITION DETAIL "NOT TO SCALE" THE 4" X 16" WALL MOUNTED WIREWAY MUST PROVIDE THREE OR FOUR FILTER PANEL MUST BE MOUNTED TO NON-CONDUCTIVE FRAMING HITACHI MEDICAL SYSTEMS, AMERICA COMPARTMENTS (AS ILLUSTRATED ABOVE) FOR RF CABLES, GRADIENT CABLES, 1959 SUMMIT COMMERCE PARK OH 44087-2371 FILTER PANEL IS TO BE MOUNTED ON BOTH SCAN AND CONTROL ROOM FINISHED WALLS (800) 800-3106 WWW.HITACHIMED.COM SIGNAL, AND POWER AND THE WIREWAY IN THE SCAN ROOM MUST FOR CABLES DROP INTO THE AN RF TIGHT SEAL IS REQUIRED ON BOTH SYSTEM WIREWAY INFORMATION BE NON-FERROMAGNETIC. (SEE OEM ADDENDUM FOR VENDOR CONTACT.) DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 9 OF 17</p><p>TYPICAL HVAC PLAN TYPICAL LIGHTING PLAN DIELECTRIC BREAK BY CONTRACTOR WAVEGUIDE WAVEGUIDE DUCT HVAC RETURN HVAC SUPPLY- TYPICAL RETURN AIR (REMOTE TEMP SENSOR 5'-0" MAX. 3-WAY DIFFUSER DUCT OUTSIDE OF RF SHIELD AWAY FROM GANTRY WITHIN 5' OF WAVEGUIDE) TEMPERATURE SENSOR GANTRY 1706 BTUH RF SHIELD INCANDESCENT LIGHTING ONLY; NO HALOGEN, SUPPLY AIR OR TRACK LIGHTING (THREE-WAY DIFFUSER) TYPICAL RECESSED CAN LIGHTS ENVIRONMENTAL NOTES 1) MRI SYSTEM ENVIRONMENTAL REQUIREMENTS: SCAN ROOM: 80°F; 45% 80% RELATIVE HUMIDITY; MAXIMUM TEMPERATURE CHANGE PER OPERATOR'S DIMMERS CANNOT BE USED 800 BTUH CAB HOUR CONSTANT, 24 HOURS/7 DAYS, CONDITIONING IS REQUIRED TYPICAL 6230 BTUH IN THE MRI SCAN ROOM OR THE CONTROL ROOM CONTROL ROOM: 70° 80°F; 45% 80% RELATIVE HUMIDITY SUPPLY AIR 2) THE TEMPERATURE IN THE MRI SCAN ROOM MUST BE MAINTAINED CONTINUOUSLY (24 HOURS/7 DAYS). IN ORDER TO ACHIEVE OPTIMAL AND CONSISTENT IMAGE QUALITY ON YOUR MRI SYSTEM, THE MAGNET TYPICAL RETURN AIR TEMPERATURE MUST REMAIN STABLE. THE NEOMAX MATERIAL USED FOR THE PERMANENT MAGNET IS VERY SENSITIVE TO TEMPERATURE VARIATION. HITACHI HAS DESIGNED A HIGHLY ACCURATE TEMPERATURE STABILIZATION SCHEME AND THIS SCHEME IS CAPABLE OF MAINTAINING THE REQUIRED TEMPERATURE WITHIN 0.1°C IF THE SCAN ROOM IS WITHIN HMSA 3) WHILE A SEPARATE AC AND HEATING UNIT IS NOT REQUIRED, HMSA RECOMMENDS A DEDICATED UNIT FOR THE MRI IMAGING SUITE CONCERNS LIGHTING NOTES MRI SCAN ROOM ONLY. A REMOTE TEMPERATURE SENSING DEVICE MUST BE LOCATED IN THE RETURN AIR DUCT, 1) THE AIRIS IS A PERMANENT MAGNET MRI SYSTEM. AS A RESULT, THE SITING 1) RECESSED CANS ARE THE LIGHT FIXTURES NEED NOT BE OUTSIDE THE RF SHIELD OF THE MRI SCAN ROOM, WITHIN 5 FEET OF THE RF THIS SHOULD BE THE REQUIREMENTS ARE MINIMAL. THERE ARE NO WATER CHILLERS OR CRYOGENS SPECIFICALLY DESIGNED FOR MRI USE. THEY MAY CONTAIN MINIMAL AMOUNTS SOLE TEMPERATURE SENSOR FOR THE HVAC UNIT SUPPLYING THE SCAN ROOM. IF A SINGLE AC NEEDED AND ELECTRICAL CONSUMPTION OF THESE SYSTEMS IS VERY LOW. OF FERROMAGNETIC MATERIALS. CARE SHOULD BE TAKEN TO ENSURE ALL UNIT IS USED TO COOL THE CONTROL AND SCAN ROOMS, IT SHOULD BE NOTED THAT WHILE THE MAJORITY OF THE 2) AUXILIARY FUNCTIONS NEEDED FOR THE EFFECTIVE OPERATION OF THE MRI SUITE WILL DECORATIVE COVERS ARE NON-FERROMAGNETIC. HEAT OUTPUT IS IN THE CONTROL ROOM, THE HVAC UNIT MUST BE CONTROLLED BY THE SENSOR IN THE REQUIRE A REVIEW OF THE EXISTING FACILITY AND ADDITIONAL MODALITIES TO BE SITED 2) NO HALOGEN, FLUORESCENT, OR TRACK LIGHTING IS TO BE USED IN THE MRI SCAN ROOM RETURN AIR DUCT. THE HVAC CONTRACTOR MUST DESIGN/BALANCE THE SYSTEM IN THE SUITE. PLUMBING WILL BE REQUIRED FOR TOILETS, FILM PROCESSORS, AND SCAN ROOM. USE ONLY INCANDESCENT LIGHTING. DUE TO THE MAGNETIC FIELD 4) A PROGRAMMABLE THERMOSTAT SHOULD BE USED AND LOCATED IN THE CONTROL ROOM WITH A LOCK BOX. SINKS THROUGHOUT THE SUITE. IN THE SCAN ROOM, LIFE-HOURS OF ALL BULBS IN THAT ROOM MAY BE REDUCED. 5) ALL AC AIR HANDLERS AND CONDENSERS MUST BE LOCATED AT LEAST 20 FEET FROM MAGNET ISOCENTER. 3) TELEPHONES AND/OR INTERCOM SYSTEM SHOULD BE MADE AVAILABLE FOR 3) DIMMERS MAY NOT BE USED IN THE MRI SCAN OR CONTROL ROOMS. A MULTI- 6) CARE MUST BE TAKEN WHEN DESIGNING THE HVAC PLAN TO ENSURE THE AIR CONDITIONING SUPPLY VENTS DO COMMUNICATION WITHIN THE MRI SUITE. AN INTERCOM BETWEEN THE TECHNICIAN SWITCH SCHEME SHOULD BE USED TO ADJUST LIGHT INTENSITIES. NOT BLOW DIRECTLY ONTO THE MAGNET. ALL HVAC DESIGN CRITERION (LIGHTS, OCCUPANCY, WINDOWS, AND THE PATIENT IS PROVIDED BY HMSA. 4) ELECTRICAL POWER FOR THE LIGHTS AND RECEPTACLES IN THE MRI SCAN AND CLIMATE, ETC.) SHOULD BE REVIEWED BY A QUALIFIED HVAC CONTRACTOR TO INSURE THAT THE ENVIRONMENTAL 4) DUE TO RISK OF STATIC SHOCK AND ISSUES REGARDING DUST, AND CONTROL ROOMS SHOULD BE DERIVED FROM THE SAME BRANCH CIRCUIT AS THE REQUIREMENTS ARE MET. OPTIONAL AND ANCILLARY EQUIPMENT (LASER IMAGER/PROCESSOR, UPS, ETC.) CARPET SHOULD NOT BE USED IN THE MRI SCAN AND CONTROL ROOMS. MRI CABINET, AND SHARE A COMMON GROUND (SEE pg. 8 of 17). THE RF SHIELDING WILL ADD SIGNIFICANTLY TO THE HEAT LOAD OF THE MRI SUITE (SEE pg. 4 of 17). IF SOMETHING OTHER THAN VCT WILL BE USED IN THE SCAN ROOM, OR IF THE THICKNESS VENDOR WILL SUPPLY AND INSTALL EMI FILTERS FOR THE SCAN ROOM LIGHTS AND 7) THE RF SHIELDING VENDOR WILL SUPPLY AND INSTALL THE RF WAVEGUIDES WHERE EACH HVAC DUCT PASSES OF THE FLOOR COVERING EXCEEDS 1/4", YOU MUST NOTIFY YOUR HMSA SITE PLANNING OUTLETS. THE LOCATION AND NUMBER OF THESE FILTERS MUST BE COORDINATED THROUGH THE RF SHIELD. LOCATIONS AND NUMBER OF THESE WAVEGUIDES MUST BE COORDINATED BETWEEN SPECIALIST AND YOUR RF VENDOR PRIOR TO INSTALLATION OF THE RF WITH THE RF CONTRACTOR. THE RF VENDOR AND HVAC CONTRACTOR. THE HVAC CONTRACTOR MUST PROVIDE FOR A DIELECTRIC CONNECTION 5) SITING REQUIREMENTS FOR THE LASER IMAGER/PROCESSOR AND UPS (IF PURCHASED) 5) DIELECTRIC CONNECTORS MUST BE USED AT THE EMI FILTER LOCATIONS IN THE BETWEEN THESE WAVEGUIDES AND HIS DUCT WORK. THESE WAVEGUIDES MAY OBSTRUCT A SMALL PERCENTAGE MUST BE TAKEN INTO CONSIDERATION WHEN PLANNING THE RF OF AIR FLOW. CONSULT YOUR RF VENDOR FOR DETAILS. 6) IF A MUSIC SYSTEM IS PLANNED FOR THE MRI SCAN ROOM, A SEPARATE FILTER MUST HITACHI MEDICAL SYSTEMS, AMERICA 8) IN THE EVENT OF INPUT POWER LOSS TO THE SYSTEM, THE MAGNET WILL BEGIN TO COOL DOWN FROM ITS OPTIMAL BE INSTALLED IN THE RF SHIELD FOR THE SPEAKER WIRES. THE SPEAKERS IN THE 1959 SUMMIT COMMERCE PARK OH 44087-2371 TEMPERATURE. FOR EVERY HOUR THE MAGNET COOLS DOWN, IT WILL TAKE TWO HOURS FOR THE MAGNET TO SCAN ROOM MUST BE LOCATED OUTSIDE THE 1 GAUSS (800) 800-3106 REACH THE PROPER TEMPERATURE AND AFTER TWELVE HOURS OF POWER OUTAGE, THE SYSTEM 7) A SINK SHOULD NOT BE LOCATED IN THE MRI SCAN ROOM. LIGHTING/ENVIRONMENTAL NOTES HEATERS ARE NO LONGER CAPABLE OF RE-HEATING THE MAGNET TO THE PROPER TEMPERATURE. THE START-UP HEATERS USED DURING INSTALLATION WILL BE REQUIRED TO RE-STABILIZE THE MAGNET. DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 10 OF 17</p><p>Y Z 12'-3" 12'-3" 0.5 GAUSS 10'-2" 0.5 GAUSS 7'-9" 1 GAUSS 1 GAUSS 6'-7" 6'-7" 16'-2" 3 GAUSS 3 GAUSS 13'-4" 5 GAUSS 5 GAUSS 16'-6" 9'-8" FLOOR TO 1 GAUSS ISOCENTER ISOCENTER ISOCENTER 0.5 GAUSS 11'-5" FLOOR TO GAUSS 5 GAUSS 3 GAUSS X Y GAUSS 6'-3" 11'-6" 14'-7" 7'-2" 7'-9' 9'-8" FIELD PLOTS 13'-6" 11'-4" "NOT TO SCALE" PLAN VIEW FRONT VIEW SIDE VIEW *SPECIAL NOTE* 3'-5" FDA REQUIREMENTS RESTRICTS PUBLIC ACCESS WITHIN THE 7'-3" 5 GAUSS FIELD UNLESS INDIVIDUALS HAVE BEEN PROPERLY SCREENED 10'-5" NOTE: THE MAGNET ISOCENTER, ONCE WILL ACT AS THE KEY FOR YOUR PLANNING. THIS REFLECTS THE LOCATION USED DURING INITIAL SITE QUALIFICATION G (TESTING), ALL SHIELDING IS REFERENCED TO THIS POINT, THE RF VENDOR POSITIONS THE SHIM PLATES ACCORDINGLY, AND LASTLY, THE RIGGER WILL REFERENCE THIS 3'-2" LOCATION WHEN SETTING YOUR SYSTEM. YOUR STRUCTURE SHOULD BE DESIGNED WITH THE ISOCENTER AS A REFERENCE POINT. 5'-3" 5'-3" 5'-2" 4'-1" HITACHI MEDICAL SYSTEMS, AMERICA 10'-6" 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 FRONT VIEW (800) 800-3106 WWW.HITACHIMED.COM SIDE VIEW LARGE FERROMAGNETIC COMPONENTS ARE GAUSS FIELDS LARGE FERROMAGNETIC COMPONENTS ARE RESTRICTED IN THE AREA NOTED. RESTRICTED IN THE AREA DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 11 OF 17</p><p>MAGNET PLACEMENT MAGNETIC FIELD PROXIMITY CHARTS THE AREA COVERED BY THE MAGNET'S FRINGE FIELDS IS OF PRIMARY CONCERN WHEN SELECTING AN MRI SITE. CONSIDERATION DEVICES AFFECTED BY MAGNETIC FIELDS MUST BE GIVEN TO THE FOLLOWING CONCERNS: 1) USE OF THE SURROUNDING AREAS, INCLUDING SPACE ABOVE, BELOW, AND ADJACENT TO THE NUCLEAR CAMERA, PET SCANNER, COLOR MONITOR, IMAGE INTENSIFIER, LINEAR 1 GAUSS 2) TYPE OF CONSTRUCTION MATERIALS USED IN THE EXISTING PARTICULARLY COLUMNS, ACCELERATOR, CT SCANNER, MRI SCANNER, ELECTRON MICROSCOPE BEAMS, JOISTS, REINFORCEMENT STEEL, AND CAST IRON PIPES. MULTI FORMAT CAMERA (LASER IMAGER), COMPUTER HARD 3 GAUSS 3) LOCATION OF MATERIALS, EQUIPMENT, AND OTHER MEDICAL DRIVE, ELECTRONIC MEMORY CARD, FERROUS CARD, WATCH 4) LOCATION OF ELEVATORS, LOADING DOCKS, PARKING LOTS, DRIVEWAYS, AND ELECTRICAL SERVICES (INCLUDING BURIED 5 GAUSS PACE MAKER, CREDIT CARD, ANALOG CELLULAR PHONE, DEFIBRILLATOR, IMPLANT UTILITIES). 5) A MAGNETIC FLUCTUATION TEST MUST BE PERFORMED AT ALL SITES TO ASSIST IN IDENTIFYING POSSIBLE SOURCES OF INTERFERENCE A SITE CANNOT RECEIVE FINAL APPROVAL BY HMSA SITE PLANNING WITHOUT THIS TEST. REQUIRED DISTANCE FROM ISOCENTER TO ITEMS 6) DO NOT SITE ANY ITEMS LISTED IN THE 20' REQUIRED DISTANCE CHART (SEE MAGNETIC FIELD PROXIMITY CHARTS). IF SITE IS THAT MAY AFFECT MAGNET AN EXISTING BUILDING, MOVE THE LISTED EQUIPMENT OUTSIDE OF THE 20' MECHANICAL/ELECTRICAL EXCLUSION ZONE. LARGE FERROUS OBJECTS OVER 400 LBS., MAJOR STRUCTURAL STEEL, FERROUS CART MAGNET SHIMMING SMALL TRUCK, HVAC AIR HANDLER, HVAC CONDENSER, WATER CHILLER, ONCE THE MRI SYSTEM IS THE MAGNETIC FIELD IS SHIMMED TO COUNTERACT THE AFFECTS OF FERROMAGNETIC 20' ELECTRICAL SUB-PANEL OVER 50 AMP, LARGE MOTOR (GAS AND ELECTRIC) OBJECTS AROUND THE STEEL EQUIPMENT, STEEL OFFICE FURNITURE (PARTICULARLY METAL FILE STORAGE RACKS), OR OTHER LARGE FERROUS OBJECTS LOCATED NEAR THE MRI SCAN ROOM SHOULD NOT BE MOVED ONCE THE MAGNET IS LARGE TRUCK OR BUS, ELEVATOR, ESCALATOR, LARGE ELECTRICAL TRANSFORMER, LARGE SHIMMED. INTRODUCING FERROMAGNETIC OBJECTS INTO THE MAGNETIC FIELD AFTER THE SYSTEM IS SHIMMED MAY CAUSE 40' ELECTRICAL PANEL (OVER 200 AMP), MAIN ELECTRICAL FEEDER LINES (INCLUDING FIELD THUS AFFECTING IMAGE QUALITY. UNDERGROUND) 200' HIGH TENSION ELECTRICAL LINES, RAILROAD TRACKS, HELICOPTER PAD MAGNET AFFECT ON LOOSE OBJECTS ATTRACTION OF LOOSE FERROMAGNETIC OBJECTS IS OF CONCERN WITHIN THE MRI SCAN ROOM. THE MAGNET'S FIELD NOTE: THE ABOVE CHARTS ARE INTENDED SOLELY AS A GUIDELINE AND NOT AS AN ALL-INCLUSIVE STRENGTH IS CAPABLE OF ACCELERATING UNRESTRAINED FERROUS OBJECTS. THE GREATER THE MASS OF THE OBJECT AND IT MAY BE POSSIBLE TO DAMPEN THE EFFECT OF SOME OF THESE INFLUENCES THE CLOSER IT GETS TO THE MAGNET, THE STRONGER THE NEAR THE MAGNET, HUMAN STRENGTH MAY NOT BE THROUGH THE USE OF MAGNETIC CONSULT WITH YOUR SITE PLANNER FOR FURTHER SUFFICIENT TO RESTRAIN FERROUS OBJECTS SUCH AS OXYGEN TANKS AND CRASH SUCH OBJECTS WILL BE DRAWN INTO (OR ONTO) THE MAGNET, POTENTIALLY CAUSING INJURY OR DEATH TO PERSONS AND DAMAGE TO THE MRI SYSTEM. SUCH OBJECTS MUST NOT BE ALLOWED IN THE SCAN ROOM. SPECIAL NON-FERROMAGNETIC EQUIPMENT SHOULD BE OBTAINED AND IDENTIFIED FOR USE IN SCAN ROOM. 4'-7 MAGNET FEET *SPECIAL NOTE* 1'-6 FDA REQUIREMENTS RESTRICTS PUBLIC ACCESS WITHIN THE 5 GAUSS FIELD UNLESS INDIVIDUALS HAVE BEEN PROPERLY SCREENED 1'-6" 8'-8 16 SHIM PLATES 4'-1 16 4'-7 16 16 THE HMSA SHIM PLATES ARE 1/8" THICK STEEL AND ARE INCORPORATED BY THE RF VENDOR INTO THE RF SHIELDED FLOOR. THE SHIM PLATES ARE DESIGNED FOR MAGNETIC FIELD HOMOGENEITY AND TO SHIELD THE MRI SYSTEM FROM FERROUS MATERIALS BELOW (SEE MAGNETIC FIELD PROXIMITY CHARTS). THEY ARE NOT INTENDED AS STRUCTURAL COMPONENTS (I.E. TO DISTRIBUTE THE WEIGHT OF THE MAGNET) NOR ARE THEY INTENDED AS LEVELING DEVICES. ASSEMBLY OF (4) HMSA SHIM PLATES ASSEMBLY OF (4) HMSA SHIM PLATES THE SHIM PLATES MUST BE SECURED IN PLACE PRIOR TO DELIVERY. IF A STRUCTURAL GLUE IS USED, IT MUST BE GIVEN SUFFICIENT BENEATH THE GANTRY (ISOMETRIC VIEW) TIME TO CURE TO PREVENT DISPLACEMENT OR BUCKLING DURING DELIVERY HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK OH 44087-2371 MAGNET PLACEMENT 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 12 OF 17</p><p>RIGGING NOTES RIGGER RESPONSIBILITIES HMSA RESPONSIBILITIES 1) SECURE INSURANCE COVERAGE FOR RIGGING, AND DELIVERY OF THE MRI SYSTEM AND 1) ASSIST THE CUSTOMER AND CONTRACTOR IN IDENTIFYING A RIGGER TO UNLOAD, UNCRATE, AND DELIVER PROVIDE THE HMSA SITE PLANNING SPECIALIST WITH A COPY OF CERTIFICATE OF INSURANCE PRIOR TO THE MRI SYSTEM. 2) ASSIST THE CUSTOMER, AND RIGGER IN IDENTIFYING THE DELIVERY PATH OF THE MRI SYSTEM. 2) DETERMINE AREA FOR UNLOADING, RIGGING, AND DELIVERING THE MRI SYSTEM COMPONENTS, INCLUDING: 3) COORDINATE WITH THE CUSTOMER, AND RIGGER THE DELIVERY DATE OF THE MRI SYSTEM. CRANE SET-UP AND CRATE UNPACKING AREA. 4) PROVIDE THE AND RIGGER WITH DELIVERY AND HANDLING PRECAUTIONS OF THE 3) INSPECT SITE PRIOR TO DELIVERY TO DETERMINE THE BEST ROUTE AND CONFIRM ADEQUATE ACCESS FOR MRI SYSTEM. DELIVERY. 5) PROVIDE THE HMSA AIRIS/AIRIS II RIGGING EQUIPMENT TO THE RIGGER. 4) PROVIDE COST ESTIMATE TO THE CUSTOMER PRIOR TO DELIVERY. 6) ASSIST THE CUSTOMER AND CONTRACTOR IN LOCATING AN ADEQUATE STORAGE SPACE FOR THE MRI 5) TAKE NECESSARY PRECAUTIONS TO PROTECT THE MRI EQUIPMENT AND THE FACILITIES (I.E. FLOORING, COMPONENTS UNTIL THE MRI INSTALLATION PROCESS IS COMPLETED. WALL COVERING, ETC.). 7) AN HMSA SERVICE REPRESENTATIVE WILL BE AT THE DELIVERY TO PROVIDE DIRECTION FOR THE UNLOADING, 6) PROVIDE NECESSARY EQUIPMENT FOR DELIVERY OF THE MRI SYSTEMS, SUCH AS: CRANE, FORKLIFT, JACKS, UNCRATING, AND PLACEMENT OF ALL HMSA MRI SYSTEM AND TOOLS. 7) APPROVE FINAL DELIVERY ROUTE. CUSTOMER/CONTRACTOR RESPONSIBILITIES 8) HMSA SHALL SEND RIGGING EQUIPMENT (SEE pg. 14 of 17) TO THE RIGGER PRIOR TO DELIVERY. THE 1) IDENTIFY, CONTRACT, AND SCHEDULE A RIGGER TO UNLOAD, UNCRATE, AND DELIVER ALL MRI SYSTEM RIGGER SHALL INSPECT THE RIGGING EQUIPMENT IMMEDIATELY UPON ARRIVAL AND NOTIFY THE HMSA YOUR HMSA SITE PLANNING SPECIALIST CAN ASSIST YOU IN IDENTIFYING AN EXPERIENCED RIGGER SITE PLANNING SPECIALIST OF ANY MISSING OR DAMAGED PARTS. IF HMSA IS NOT NOTIFIED PRIOR TO IN YOUR AREA. THE DELIVERY DATE WILL BE A COORDINATED EFFORT BETWEEN THE CUSTOMER, DELIVERY, MISSING OR DAMAGED RIGGING EQUIPMENT WILL BE REPLACED AT THE RIGGER'S HMSA, AND THE RIGGER. THE DELIVERY DATE IS SET BASED ON SITE MRI SYSTEM AVAILABILITY, 9) UNLOAD ALL THE MRI EQUIPMENT, AND DELIVER IT INSIDE AS DIRECTED BY HMSA SERVICE RIGGER AVAILABILITY, AND ADMINISTRATIVE REPRESENTATIVE ON-SITE. 2) PROVIDE A TYPICAL 8'-0"w X 7'-0"h OPENING AND DELIVERY PATH FOR THE MAGNET. 10) REMOVE THE FRONT LIFTING JIG FROM THE MAGNET ONLY AFTER THE MAGNET IS SET INTO PLACE. 3) ENSURE THE AREA REQUESTED BY THE RIGGER FOR DELIVERY TRUCKS, CRANE, AND WORK SPACE IS AVAILABLE 11) DISPOSE OF ALL PACKING AND CRATING MATERIALS (INCLUDING STEEL SHIELD PLATES AND CRIBBING IN AND FREE OF OBSTRUCTIONS. THE OVERSEAS CONTAINER, SHOCK ABSORBERS, PALLETS, AND STRAPPING CABLES), AS DIRECTED BY 4) ENSURE THE DELIVERY PATH FOR THE MAGNET AND SYSTEM COMPONENTS IS CLEAR AND FREE OF THE HMSA SERVICE REPRESENTATIVE ONLY THE RIGGING EQUIPMENT SENT BY HMSA, THE FRONT LIFTING 5) HAVE A FACILITY REPRESENTATIVE AVAILABLE AT THE TIME OF DELIVERY. JIG AND ITS BOLTS, AND FOUR (4) M-36 EYEBOLTS ORIGINALLY ATTACHED TO THE MAGNET ARE TO BE SAVED. 6) MANAGE ALL COSTS ASSOCIATED WITH TRANSPORTING AND RIGGING THE MRI SYSTEM. 12) MAKE SURE ALL RIGGING EQUIPMENT SENT BY HMSA IS RETURNED TO ITS SHIPPING CRATE. THE RIGGER 7) SIGN SHIPPING DOCUMENTS UPON DELIVERY. IS RESPONSIBLE FOR THE RIGGING EQUIPMENT UNTIL IT IS RETURNED TO HMSA. 8) REMOVE OR PROTECT ANY ITEMS ALONG THE MAGNET DELIVERY PATH THAT MAY BE AFFECTED BY THE MAGNETIC 13) CONTACT THE HMSA SITE PLANNING SPECIALIST AFTER DELIVERY TO ARRANGE FOR PICK-UP OF THE FIELD (I.E. MAGNETIC STORAGE UNITS, COMPUTERS, DISKETTES, ETC.). RIGGING EQUIPMENT AND THE FRONT LIFTING JIG. 9) SECURE INSURANCE COVERAGE, EFFECTIVE UPON DELIVERY OF YOUR SYSTEM. 10) PROVIDE A SECURE STORAGE AREA NEAR THE MRI CONTROL ROOM (MINIMUM 10' 10') FOR MRI SYSTEM MATERIALS NEEDED FOR DELIVERY (PROVIDED BY RIGGER) COMPONENTS UNTIL THE MRI INSTALLATION PROCESS IS A CRANE (SIZED BY THE RIGGER) FOR UNLOADING THE MAGNET FROM THE OVERSEAS CONTAINER, ROLLERS 11) IF A SECURABLE STORAGE AREA IS NOT AVAILABLE, NORTH AMERICAN VAN LINES WILL STORE THE UNIT FOR A OR AIR BAGS TO MOVE THE MAGNET INTO PLACE, FORKLIFT (SIZED BY THE RIGGER) TO UNLOAD MRI COMPONENTS, SECOND THE CUSTOMER IS RESPONSIBLE FOR THE ADDITIONAL COSTS. JACKS, STEEL PLATES AND/OR PLYWOOD AS NEEDED TO PROTECT EXISTING FLOORING, PRYBARS TO UNCRATE MRI COMPONENTS, #17 & #22 METRIC ALLEN WRENCHES, AND HMSA AIRIS/AIRIS II RIGGING EQUIPMENT (PROVIDED BY HMSA). THE RIGGER WILL PROVIDE ANY ADDITIONAL EQUIPMENT (SUCH AS CRANE MAT AND ADDITIONAL TOOLS) AS NEEDED. HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 (800) 800-3106 WWW.HITACHIMED.COM RIGGING PROCEDURE DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 13 OF 17</p><p>AIRIS II CRATED COMPONENTS HMSA AIRIS/AIRIS II RIGGING EQUIPMENT THIS RIGGING EQUIPMENT WILL BE SENT BY HMSA TO THE RIGGER PRIOR TO DELIVERY. UPON THE RIGGER SHALL AIRIS II COMPONENT LIST INVENTORY THE EQUIPMENT AND IMMEDIATELY REPORT ANY MISSING OR DAMAGED PARTS. AFTER THE MRI SYSTEM DELIVERY, COMPONENT LENGTH WIDTH HEIGHT NET GROSS THE RIGGER SHALL RE-CRATE THE RIGGING EQUIPMENT FOR PICK-UP BY HMSA. ANY LOST OR DAMAGED ITEMS, NOT REPORTED CRATE NUMBER (CRATED) (CRATED) (CRATED) WEIGHT (lbs) WEIGHT (lbs) PRIOR TO DELIVERY, WILL BE REPLACED AT THE RIGGER'S EXPENSE. SEE BELOW FOR INVENTORY OF HMSA RIGGING GEAR. 1. GANTRY 8'-6" 6'-0" 5'-11" 34,615 35,935 2. GANTRY COVERS 8'-11" 5'-9" 7'-0" * * 3. PARTS 7'-3" 3'-6" 3'-6" * * 4. PATIENT TABLE 8'-3" 3'-2" 2'-7" 882 1,257 5. MRI CABINET 3'-7" 3'-0" 6'-8" 1213 1,169 (2) LARGE (16) (2) LARGE "JACK BOSSES" (2) SMALL "JACK BOSSES" M-64 EYEBOLTS HEX BOLTS 6. FILTER PANEL 4'-3" 2'-10" 1'-9" 133 221 7. CABLES 4'-2" 2'-10" 2'-3" NON-FERROUS NOTE: DUE TO THE WEIGHT OF THE MRI SYSTEM COMPONENTS, A RIGGER IS NECESSARY FOR UNLOADING M-64 LARGE AND PLACING THE MRI SYSTEM. UNLESS OTHERWISE SPECIFIED IN THE CONTRACT, HIRING A RIGGER IS 9'-0" EYEBOLTS THE RESPONSIBILITY OF THE CUSTOMER, NOT HMSA. THE SITE PLANNER CAN ASSIST THE CUSTOMER M-64 LARGE FRONT LIFTING JIG EYEBOLT LOCATIONS LOCATION AND/OR CONTRACTOR IN IDENTIFYING A RIGGER IN THE AREA THAT HAS EXPERIENCE WITH THE MRI SYSTEM. ONCE A RIGGER IS HIRED, NOTIFY THE HMSA SITE PLANNER so THAT HE CAN ASSIST IN COORDINATION OF THE DELIVERY AND SUPPLY THE RIGGER WITH THE NECESSARY RIGGING EQUIPMENT. A CERTIFICATE OF INSURANCE MUST BE ON FILE WITH HMSA SITE PLANNING DEPARTMENT PRIOR TO DELIVERY. SMALL JACK BOSS -SMALL JACK BOSS 5'-11" LARGE 6'-0" JACK BOSS LARGE LARGE AIRIS II COMPONENT TRUCK LOADS AND REMOVAL JACK BOSS JACK BOSS LARGE JACK BOSS 6 5 3 8'-6" FRONT LIFTING 2 JIG TOP VIEW FRONT VIEW 7 4 1) THE FRONT LIFTING JIG MUST REMAIN IN PLACE UNTIL THE MAGNET IS IN ITS FINAL POSITION. THE RIGGER WILL REMOVE THE FRONT LIFTING JIG AS INSTRUCTED BY THE HMSA SERVICE REPRESENTATIVE ON SITE. HMSA WILL MAKE ARRANGEMENTS WITH THE RIGGER TO PICK UP ALL HARDWARE ATTACHED TO THE MAGNET UPON DELIVERY. THESE INCLUDE THE FOLLOWING: 20' OVERSEAS DRY CONTAINER ONE (1) LIFTING JIG, ALL LIFTING JIG BOLTS, AND FOUR (4) M-36 EYEBOLTS. TOP VIEW 2) THE RIGGER IS RESPONSIBLE FOR RETURNING ALL HMSA RIGGING EQUIPMENT TO ITS CRATE AND PREPARING IT FOR PICK UP. 15° LOST OR DAMAGED RIGGING EQUIPMENT MUST BE REPORTED IMMEDIATELY TO THE HMSA SITE PLANNING SPECIALIST. 3) THE ACCESS AND DELIVERY ROUTE FOR THE MAGNET MUST ALLOW FOR ALL RIGGING MATERIALS AND SAFE WORKING NOTE 1: SYSTEM PRECAUTIONS: THIS IS A PERMANENT MAGNET. THE MAGNET IS FULLY ACTIVE WHEN IT ARRIVES ON THE TRUCK AND CANNOT BE TURNED OFF. THE MAGNETIC FIELD FOR THE HMSA SYSTEM IS THE STRONGEST AVAILABLE PERMANENT MAGNET ON THE MARKET AND IT WILL ATTRACT FERROUS MATERIALS SUCH AS METAL PLATES, JACKS, TOOLS, ETC. THE MAGNETIC FIELD WILL ALSO 1 AFFECT OBJECTS LIKE WATCHES, PAGERS, CELL PHONES, CAMERAS, CREDIT CARDS, AND RECORDING DEVICES (SEE pg. 12 of 17). HITACHI MEDICAL SYSTEMS, AMERICA NOTE 2: NORTH AMERICAN VAN LINES WILL STORE THE ANCILLARY 1959 SUMMIT COMMERCE PARK OH 44087-2371 20' OPEN TOP OVERSEAS DRY CONTAINER ANGLE OF THE GANTRY MUST NOT EXCEED (800) 800-3106 COMPONENTS OFF-SITE IF ON-SITE STORAGE IS NOT TOP VIEW 15 DEGREES AT ANY TIME DURING DELIVERY COST OF SECOND DELIVERY IS THE RESPONSIBILITY OF THE CUSTOMER. RIGGING DETAILS DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 14 OF 17</p><p>SCAN ROOM 6" 8" MINIMUM SERVICE 6' ACCESS AREA 21 SPEAKER FRONT VIEW SPEAKER SIDE VIEW 2 SPEAKER -7" THE SPEAKER IS WALL MOUNTED IN THE SCAN ROOM. A 1/2" WIREMOLD FROM THE WIREWAY CONNECTS TO THE BOTTOM EQUAL FILTER PANEL BOTTOM VIEW OF THE UNIT. THE UNIT IS MOUNTED WITH A SINGLE SCREW. 27" 7'-6" 14'-0" 6' 3'-3" 10'-4" 6" CLEAR MIN. 13.5" 23-5/8" ROUGH RF SEAL BOTH SIDES 3'-3' 1'-5" RF SEAL BOTH SIDES OPENING 1/2" AVAILABLE) 16.5" MIN. REQUIRED SERVICE EQUAL (1 1/2" AVAILABLE) MAG. COMP. UNIT TOP VIEW 3'-0" ACCESS AREA 7'-6" RECOMMENDED COMPENSATION 3'-3" COIL 28" 4" X 16" WALL MOUNTED WIREWAY -6" RF FLOOR -6" FLUSH WITH BOTTOM OF FILTER PANEL MAG. COMP. UNIT FRONT VIEW TO ROUGH OPN'G 6'-6" LOCATED ON FLOOR IN THE LEFT CLEAR MIN. BACK CORNER OF SCAN ROOM G 3'-0" FILTER PANEL FRONT VIEW 18'-2" CLEAR MIN. PLAN VIEW FINISHED WALLS MAGNETIC COMPENSATION UNIT NON-CONDUCTIVE THE MAGNETIC FIELD COMPENSATION UNIT IS REQUIRED WHEN OUTSIDE INTERFERENCE FRAMING HAS BEEN IDENTIFIED. IT IS COMPRISED OF (3) COMPONENTS: THE COMPENSATION UNIT RF (SEE ABOVE), THE COMPENSATION COIL (SEE ABOVE), AND THE PROBE (NOT SHOWN). IT G SEAL FILTER PANEL IS DESIGNED TO STABILIZE THE MAGNET'S FIELD WHEN SUBJECTED TO THIS OUTSIDE 35-7/16" ROUGH 44" INTERFERENCE (SEE pg. 12 of 17). IF IT IS DETERMINED THROUGH HMSA TESTING THAT A OPENING FILTER PANEL COMPENSATION UNIT IS REQUIRED, CONTACT YOUR HMSA SALESPERSON FOR DETAILS. COVER - THE UNIT MUST BE LOCATED IN THE SCAN ROOM, OUTSIDE OF THE 5 GAUSS FIELD. SEAL - THE COIL MUST BE CENTERED OVER THE MAGNET ISOCENTER WITH A 6" CLEAR RF WALL SERVICE AREA FOR ADJUSTMENTS (7'-6" X 7'-6" TOTAL AREA). ITS WEIGHT IS 28' 6'-5" 16" APPROXIMATELY 40 POUNDS AND IS MOUNTED AT A MAXIMUM OF 8'-2" A/FF. SUPPORT IS RF FLOOR TO ROUGH OPN'G WIREWAY BY YOUR SUSPENDED CEILING. THE LOCATION OF THE COIL MUST BE CONSIDERED 3'-2" SCAN ROOM WHEN FINALIZING YOUR CEILING PLAN. 3'-0" 4" TYP. CONTROL ROOM RF FLOOR AIR GAP THE PROBE LOCATION CANNOT BE DETERMINED UNTIL IT WILL BE WALL WALL CROSS-SECTION MOUNTED IN PROXIMITY OF THE SOURCE OF WIREMOLD (1/2"), BY REQUIRED FOR CABLE CONNECTIONS BETWEEN ALL SIDE VIEW FILTER PANEL SIDE VIEW COMPONENTS AND THE WALL WIREWAY. AIRIS II GANTRY AND PATIENT TABLE WALL MOUNTED FILTER PANEL HITACHI MEDICAL SYSTEMS, AMERICA MINIMUM SERVICE CLEARANCES REFLECTED IN THE PLAN ABOVE. RF VENDOR WILL MOUNT THE FILTER PANEL. COORDINATE WITH 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 CONTACT SITE PLANNING IMMEDIATELY WHEN SERVICE AREAS VENDOR REGARDING ASSEMBLY (800) 800-3106 WWW.HITACHIMED.COM MAY BE SEE pg. 9 of 17 FOR MOUNTING REQUIREMENTS - THE MAGNET ISOCENTER MUST BE REFERENCED TO MAINTAIN COMPONENTS - SCAN ROOM CONSISTENCY FOR PLANNING PURPOSES - SEE pg. 4 of 17 FOR GANTRY AND PATIENT TABLE DIMENSIONS DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 15 OF 17</p><p>CONTROL ROOM AS WITH MOST IMAGING MODALITIES, YOU MAY WISH TO CONNECT YOUR AIRIS TO AN EXISTING NETWORK OR DEVELOP A NEW NETWORK TO BETTER MEET YOUR FACILITIES NEEDS. THIS MIGHT UTILIZE THE HMSA AVIA WORKSTATION AS THE CENTRAL 4'-0" 5'-4" COMPONENT TO COMMUNICATE WITH OTHER MODALITIES, OR YOU MAY WANT TO CONNECT WITH REMOTE VIEWING STATIONS OR ADDITIONAL PRINTERS/PLOTTERS. YOUR NETWORK ADMINISTRATOR SHOULD REVIEW YOUR FACILITIES REQUIREMENTS AND 36.5" DEVELOP A CONNECTIVITY YOUR NETWORK ADMINISTRATOR MUST THEN CONTACT HMSA TECHNICAL SUPPORT 2'-6" 09 REGARDING YOUR SPECIFIC STANDARDS AND CONFIGURATION TO DETERMINE COMPATIBILITY WITH OUR EQUIPMENT (4) CASTERS (4) LEVELING JACKS 2.5"DIA. X 1.5"W 31 2.5"DIA. W/.75" BOLT 23.7" LASER IMAGER TOP VIEW 6'-7" 35.3" A LASER IMAGER IS A COMMON COMPONENT THAT SHOULD BE PART OF YOUR INSTALLATION DESIGN. MANY MODELS OF LASER IMAGERS ARE AVAILABLE. WHEN CHOOSING YOUR LASER IMAGER, ENSURE THE UNIT HAS: + 1'-0" 1'-0' 1) SUFFICIENT MEMORY TO OPERATE WELL WITH THE HITACHI EQUIPMENT AND OTHER MODALITIES THAT WILL SERVICE AREA SHARE THE LASER IMAGER'S SERVICE. THE HITACHI MR SYSTEM REQUIRES A MINIMUM OF 64MB OF IMAGER 3'-0" MEMORY FOR OPTIMAL 2) A DIGITAL OR DICOM 3.0 (OPTION) INTERFACE CAPABILITY, INCLUDING HARDWARE AND SOFTWARE. 3) A CABLE CONNECTION RUN (WIREWAY) OR NETWORK JACK, NEAR THE HITACHI LASER IMAGER INTERFACE, BOTTOM VIEW 2'-6" DESIGNED BY YOUR ARCHITECT, AVAILABLE AT THE TIME OF *SEE YOUR HMSA SALES REPRESENTATIVE FOR MORE INFORMATION ON HITACHI'S LASER IMAGER. 3'-4" 2'-7" POWER CONDITIONING FRONT VIEW SIDE VIEW IF YOU SUSPECT YOUR SITE HAS POWER PROBLEMS AND YOU WANT TO ENSURE OPERATION THROUGHOUT THESE PROBLEMS, HMSA CAN PROVIDE A UPS SYSTEM FOR THE AIRIS II SYSTEM. OPERATOR'S WORKSTATION THE ABOVE COMPONENTS CONSTITUTE THE EXHAUST FANS THE OPTIONAL UPS, IF PURCHASED FROM HMSA, IS SIZED FOR THE AIRIS II SYSTEM ONLY. PRIOR TO ORDERING ANY TYPE OF POWER OPERATOR'S WHILE A TABLE CONDITIONING DEVICE, YOU SHOULD FIRST DETERMINE WHAT TYPE OF SERVICE YOU DESIRE AND WHAT TYPE OF POWER IS PROVIDED BY HMSA, THE DESIGN OF AN CONCERNS ARE PRESENT AT YOUR SITE. 5'-10" ADJACENT WORK COUNTER WITH STORAGE POWER CONCERNS: SHOULD ALSO BE CONSIDERED (SEE BELOW). 1) WHAT IS THE FREQUENCY OF YOUR POWER DISRUPTIONS? IF THESE DISRUPTIONS OCCUR WITH MINIMAL FREQUENCY, YOU MAY NOT NEED ANY POWER CONDITIONING 2) DO THESE DISRUPTIONS TAKE THE FORM OF SAGS, SURGES, OR VOLTAGE FLUCTUATIONS? COUNTER BY OTHERS CLEARANCE A VOLTAGE REGULATOR OR POWER CONDITIONER MAY BE ALL THAT IS NEEDED. (OPTIONAL) (MIN) 3) ARE THESE TOTAL BLACKOUTS? CABLES - A UPS WOULD BE REQUIRED, BUT THIS WOULD ONLY POWER THE SYSTEM FOR A MINIMAL TIME. ADDITIONAL CONSIDERATIONS NEED TO BE REVIEWED FOR THE ENTIRE SUITE. 4) HAS YOUR ELECTRICIAN MONITORED YOUR POWER WITH A DRANITZ METER TO IDENTIFY THE CONCERNS? FRONT VIEW SIDE VIEW THIS IS THE ONLY WAY YOU WILL GET A TRUE READING OF YOUR POWER CONCERNS. 1'-6" DESIRED 4'-0" MIN. 2'-6" MRI CABINET 1) DO YOU WISH TO SAVE THE DATA YOU HAVE ACQUIRED? THE SYSTEM SHOULD RETAIN EXISTING DATA BUT A SMALL UPS WOULD ALLOW FOR A STANDARD SHUT DOWN. - 208 V; 50 AMP (60 AMP IF UPS IS INSTALLED) 2) DO YOU WISH TO FINISH THE PRESENT SCAN? MRI CABINET MUST BE LOCATED OUTSIDE THE 1 GAUSS FIELD - THE OPTIONAL HMSA UPS CAN PROVIDE POWER TO THE SYSTEM FOR APPROXIMATELY 20 MINUTES. ADEQUATE SPACE MUST BE PROVIDED FOR SERVICE AND FRONT VIEW SIDE VIEW HOWEVER, THE OTHER SUITE FUNCTIONS (LIGHTS, OUTLETS, HVAC) MAY ALSO REQUIRE BACK-UP POWER. AN COOLING ADDITIONAL UPS MAY BE REQUIRED OR YOU MAY WANT YOUR DESIGN TEAM TO DESIGN A LARGER UNIT. 3) DO YOU WISH TO MAINTAIN OPERATIONS? NETWORK THE HMSA SUPPLIED UPS WOULD HANDLE YOUR SYSTEM FOR APPROXIMATELY 20 MINUTES. THE UPS, AS WELL AS PERIPHERAL EQUIPMENT ASSOCIATED WITH MOST MEDICAL IMAGING INSTALLATIONS CAN GREATLY AFFECT THE ACCEPTABILITY OF HITACHI MEDICAL SYSTEMS, AMERICA THE OTHER SUITE FUNCTIONS, WOULD NEED TO BE 1959 SUMMIT COMMERCE PARK TWINSBURG, OH 44087-2371 YOUR PLANS AND THE SYSTEM OPERATION. YOUR ARCHITECT MUST ENSURE PROVISIONS ARE MADE FOR ALL PERIPHERAL BACKED UP BY AN EMERGENCY GENERATOR. (800) 800-3106 WWW.HITACHIMED.COM EQUIPMENT, INCLUSIVE OF SPACE REQUIREMENTS, POWER, HEAT LOAD, AND THE SELECTION OF THESE COMPONENTS CONTROL ROOM COMPONENTS AND THEIR PLACEMENT MUST BE REVIEWED WITH HMSA SITE PLANNING PRIOR TO FINALIZATION OF YOUR DESIGN. DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS SHT 16 OF 17</p><p>AIRIS II SITE SELECTION SURVEY AIRIS II SITE CHECKLIST FOR INSTALLATION FOR MULTIPLE LOCATIONS? SITE LOCATION: FSE: LIST NUMBER OF LOCATIONS: S/N: ADDRESS: VM #: AIRIS II LOCATION #1 SITE NAME: THE FOLLOWING ITEMS MUST BE COMPLETED BEFORE INSTALLATION CAN BEGIN. TYPE OF CONSTRUCTION: EXISTING NEW ADD ON MODULAR CUSTOMER RESPONSIBILITIES: YES NO CONSTRUCTION: 1. CLEAR AVAILABLE SPACE (15'-0" WIDE X 20'-0" LONG) FOR SCAN ROOM ONLY? SCAN ROOM AND CONTROL ROOM CONSTRUCTION COMPLETE 2. ADJACENT SPACE AVAILABLE FOR CONTROL AND EQUIPMENT ROOMS? WIREWAYS MUST BE INSTALLED IN SCAN ROOM AND CONTROL ROOM AS SPECIFIED BY HITACHI RF FILTER PANEL MUST BE ACCESS OPENING CLOSED, AND ROOM TESTED 3. SITE HAS AT LEAST 8'-0" OPEN CLEARANCE ABOVE FINISHED FLOOR? ALL ROOMS MUST BE CLEAN OF DUST AND DEBRIS 4. SLAB ON-GRADE CONSTRUCTION FOR SCAN ROOM? RF ROOMS MUST BE FLOATING IN RELATION TO GROUND (NOT SHORTED TO EARTH GROUND) ROOM FINISHES COMPLETE WALLS 5. DOES SITE CONTROL SPACE IF BASEMENT IS LOCATED BELOW? CEILING 6. MULTISTORY BUILDING WITH FLOORS ABOVE SCAN ROOM? FLOORING ELECTRICAL: 7. DOES SITE CONTROL SPACE IF FLOORS ABOVE? 220 VAC; 50 AMP (60 AMP REQUIRED WITH A UPS); SINGLE PHASE POWER INSTALLED AND 8. ANY LARGE (400 POUNDS) FERROMAGNETIC STRUCTURAL MEMBERS LOCATED WIRED TO DISCONNECT WITH PIGTAIL WITHIN 10'-0" OF MAGNET CENTER? CONTROL AND SCAN ROOM LIGHTING AND OUTLETS OPERATIONAL ROOM MUST HAVE A SINGLE POINT GROUND 9. SMALL VEHICLES (CARS, SUV'S), HVAC UNITS WITHIN 20'-0" OF MAGNET CENTER? ENVIRONMENTAL: 10. LARGE VEHICLES (TRUCKS, BUSES) AND ELEVATORS WITHIN 40'-0" OF MAGNET CENTER? HVAC OPERATIONAL IN SCAN ROOM AND CONTROL ROOM - MUST MEET SYSTEM 11. TRAINS, SUBWAYS, MAJOR HIGHWAYS, OR HIGH VOLTAGE POWER LINES IN THE IMMEDIATE AREA? SPECIFICATION: 80°F, 24 HOURS/DAY, 7 DAYS/WEEK 12. ELECTRICAL SERVICE OF 50 AMPS OR GREATER (TRANSFORMERS, PANELS, POWER LINES) FACILITIES: WITHIN 40'-0" OF MAGNET CENTER? HMSA MUST HAVE 24-HOUR ACCESS TO FACILITY TABLE, CHAIRS, AND RESTROOM FACILITIES AVAILABLE IN WORK AREA. 13. ANY VIBRATION INDUCING SYSTEMS (MOTORS, PUMPS, HVAC UNITS) WITHIN THE AREA? ALL OF THE ITEMS LISTED ABOVE HAVE BEEN CHECKED AND ARE COMPLETE: 14. ANY EXISTING OR PROPOSED MEDICAL IMAGING EQUIPMENT IN THE AREA? 15. IS EQUIPMENT DELIVERY ACCESS AVAILABLE WITHIN THE FACILITY? DATE: HMSA FSE NOTE: THIS CHECKLIST MUST BE PRESENTED TO THE DISTRICT MANAGER OR GROUP LEADER AT TIME OF INSTALL AUTHORIZATION TO START INSTALL WITHOUT DISTRICT MANAGER OR TWO GROUP LEADERS GL #1 GL #2 HITACHI MEDICAL SYSTEMS, AMERICA 1959 SUMMIT COMMERCE PARK OH 44087-2371 (800) 800-3106 PRELIMINARY AND FINAL CHECKLISTS DATE: 1/03/00 REVISION: 03/08/05 AIRIS STANDARD DETAILS SHT 17 OF 17</p><p>OEM PRODUCTS AND SERVICES OEM PRODUCT AND SERVICE VENDORS ARE KEY TO YOUR SUITE PREPARATION. THEY SHOULD BE INVOLVED EARLY IN THE PLANNING PROCESS WHEN POSSIBLE. WHILE VENDORS LISTED BELOW HAVE DEMONSTRATED RF VENDORS: STRUCTURAL SLAB ISOLATION MATERIAL ACCEPTABLE CUSTOMER SERVICE IN THE PAST, THIS IN NO WAY IS TO BE CONSTRUED AS AN ENDORSEMENT OR BRADEN SHIELDING TULSA, OK UNISORB A GUARANTEE OF FUTURE SERVICE. THIS LISTING IS PROVIDED AS A CONVENIENCE ONLY AND IS BY NO MEANS PHONE: (918) 624-2888 PHONE: (517) 764-6060 THE CUSTOMER IS RESPONSIBLE FOR CHOOSING THEIR VENDORS. CONTACT: HAROLD JIMISON CONTACT: TIM SPAHR UPON IDENTIFYING YOUR VENDORS, PROVIDE CONTACT INFORMATION TO HMSA SITE GAVEN INC. PITTSBURGH, PA WIREWAYS PHONE: (412) 821-8312 WIREMOLD/WALKER CONTACT: RICHARD PAHLER, JR. RIGGING COMPANIES: PHONE: (304) 375-1174 CONTACT: MIKE MARLOW NORCAL RIGGERS CA HWP RIGGING ST. LOUIS, MO GLOBAL PARTNERS IN SHIELDING PHONE: (510) 357-6400 PHONE: (314) 865-0100 PASSAIC, NJ MISC. VENDORS STEVE OWEN CONTACT: ROBERT DEUTSCH PHONE: (973) 574-9077 SAFETY FERROGUARD CONTACT: MARK HOLDER LINDGREN RF ENCLOSURES J.C. DUGGAN, INC. NY HEAVY COMPANY LINCOLN, NE PHONE: (718) 384-3260 PHONE: (800) 769-6474 GLENDALE HEIGHTS, IL IMEDCO USA NOBLESVILLE, IN CONTACT: JOHN CEREGHINO CONTACT: TIM GEIS PHONE: (630) 307-7200 PHONE: (317) 770-0655 CONTACT: ERIC KAAD CONTACT: MICHAEL KRANCHON ALL AMERICAN HOUSTON, TX NIEMAN & COMPANY RIGGING AND CRANE AIRIS II COLORS PHONE: (713) 433-6359 SERVICES INC. FL LINDGREN RF ENCLOSURES, INC. THE FOLLOWING ARE SHERWIN WILLIAMS PAINT CODES CONTACT: JOHN CARSON PHONE: (904) 783-8464 GLENDALE HEIGHTS, IL WHICH MATCH THE AIRIS COLORS: CONTACT: DARIN NIEMAN PHONE: (630) 307-7200 HAMMERNET VEGA WHITE J22 XXW 8907 4337 ABLE MACHINERY MOVERS, INC. TX CONTACT: BRET WATSON HAMMERNET REGAL GREY #5 J22 XXA 8908 4337 (800) 856-1914 ADVANCED RIGGING AND LOGISTICS HAMMERNET EGGSHELL BLUE J22 XXW 8909 4337 CONTACT: MIKE ALIANELL CHINO, CA SHIELDING DYNAMICS INC. TX PHONE: (877) 688-2275 PHONE: (281) 290-8351 DIAMOND RIGGING CORPORATION BATAVIA, IL CONTACT: STEVE GEBHART NOTE: THE ABOVE REPRESENTS A PARTIAL LISTING OF CONTACT: DAVID WERNICK PHONE: (630) 879-6500 ACCEPTABLE VENDORS LOCATED ACROSS THE CONTACT: MAX MAYER MEDICAL EQUIPMENT CONTRACTORS COUNTRY. CONTACT SITE PLANNING FOR ADDITIONAL SRG TULSA OK WESTCHESTER, PA SELECTIONS. PHONE: (918) 663-1985 B & B RIGGING NORTH LARGO, FL (610) 321-0694 CONTACT: MIKE LAHITA PHONE: (727) 539-6691 CONTACT: JOE SWEENEY CONTACT: RENDER BRITTON UNIVERSAL SHIELDING DEER PARK, NY PLITT CRANE AND RIGGING SERVICE PHONE: (631) 667-7900 BELGER RIGGING KANSAS CITY, MO TX IRWIN NEWMAN PHONE: (816) 472-0000 PHONE: (956) 831-4278 CONTACT: BUD BOWLES CONTACT: WALTER E. PLITT RIOPELLE DEVELOPMENT CA PHONE: (760) 439-6344 J.V. RIGGING NEW KENSINGTON, PA SOUTHWEST RIGGING PHOENIX, AZ CONTACT: TIM RIOPELLE PHONE: (724) 339-8900 PHONE: (602) 278-6281 CONTACT: JOE VAGLIA HITACHI MEDICAL SYSTEMS, AMERICA CONTACT: JIM JOBE MRI CORPORATION IRVINE, CA 1959 SUMMIT COMMERCE PARK OH 44087-2371 PHONE: (714) 668-4244 (800) 800-3106 ATLAS INDUSTRIAL CONTRACTORS COLUMBUS, OH BOB RUMNEY OEM ADDENDUM PHONE: (800) 562-8322 CONTACT: RAY BUDD DATE: 01/03/00 REVISION: 03/08/05 AIRIS II STANDARD DETAILS OEM</p>