Abstract Aim The aim of this work was to design an accurate 3D digital model of the humerus and rotator cuff muscles. This model was then used to study strain distribution in humeral tubercles according to bone density. Materials and methods The geometry of bone and muscle structures was reproduced using SURFDRIV- ER� software, based on anatomical sections, CT scans and MRI images from the Visible Human Project image library. The contours were transferred to PATRAN� software to rebuild volumes and mesh them. Calculations of strains and their distribution were performed using NASTRAN� software. All the elements were considered to be isotropes. Results The study of the distribution of stress magnitude according to the type of bone modeled, shows that some stresses in cortical bone are greater than those in cancellous bone and are also greater in old bone, implying more deformation in old bone at constant force. This study also shows that stresses do not penetrate deeply into cancellous tissue. Conclusion Observing the simulation results led understanding of the pathology of certain fractures of the proximal end of the humerus. This study also helped explain why certain types of osteosynthesis fail due to tubercles reconstruction failures. Keywords Shoulder Æ Humerus Æ Finite element analysis Æ Stress distribution Æ Biomechanics Introduction Many physical or computer models of the scapulohu- meral joint have been developed to improve under- standing of its function. Some were developed to understand and analyze muscle action developed during certain movements of the upper limb (inverse dynamic theories) [4, 13, 15, 23, 24, 26]. Others, based on the concept of deformable bodies, were proposed and used to calculate the distribution of bone stresses [3, 10, 16, 20, 22]. Anatomically, four rotator cuff muscles (supraspin- atus, infraspinatus, subscapularis, teres minor) insert into the humeral tubercles, where stress is concen- trated during abduction movements. The purpose of this work was to develop a 3D digital model of the proximal end of the humerus and the rotator cuff muscles, in order to analyze the effect of bone tissue quality on strain distribution to the proximal end of the humerus owing to the physiological action of rotator cuff tendons during an abduction movement. Materials and methods Obtaining anatomical shapes in 3D SURFDRIVER� (SURFDRIVER, Kailua, HI, USA) software was used for acquisition of the external Ph. Clavert (&) Æ J. F. Kempf Æ J. L. Kahn Biomechanical Laboratory of the GEBOAS, Institute of Normal Anatomy, Faculty of Medicine, 4 rue Kirschleger, 67085 Strasbourg Cedex, France e-mail: philippe.clavert@chru-strasbourg.fr M. Zerah Æ J. Krier Æ P. Mille Laboratoire d’Ingénierie des Surfaces de Strasbourg Groupe LISS du LGECO, INSA Strasbourg, 24 boulevard de la victoire, 67084 Strasbourg Cedex, France Surg Radiol Anat (2006) 28:581–587 DOI 10.1007/s00276-006-0140-x 123 ORIGINAL ARTICLE Finite element analysis of the strain distribution in the humeral head tubercles during abduction: comparison of young and osteoporotic bone Ph. Clavert Æ M. Zerah Æ J. Krier Æ P. Mille Æ J. F. Kempf Æ J. L. Kahn Received: 4 April 2006 / Accepted: 3 July 2006 / Published online: 26 August 2006 � Springer-Verlag 2006 contours of the different elements of the bone (cortical bone, cancellous bone, subchondral bone) and articular cartilage found in the proximal third of the humerus (Fig. 1). This was done using 153 anatomical and CT scan sections from the male specimen of the Visible Human Project image library (National Library of Medicine, Rockville Pike, USA). They were millimet- ric coronal sections in 1 mm steps. Since the automatic outlining abilities of currently available software are not suitable for the complexity of the anatomical structures studied, this outlining was then accurately defined on each CT slice (digitization error lower than 0.7 mm, i.e. 2 pixels). Using the same process, the muscle bodies of the subscapular, supraspinatus (Fig. 2), infraspinatus and teres minor were outlined. Finally, the insertion areas of rotator cuff muscles and their tendons were defined by comparison with anatomical sections and MRI images. Meshing The contours obtained were transferred to MSC PATRAN v7.5 (MCS Software Corporation, Santa Ana, CA, USA) to be assembled individually and thus reconstruct the volume of each element before mesh- ing them. Meshing or discretizing all the elements constituting the reconstructed humerus, was carried out with 8 node brick type isoparametric three- dimensional elements or HEX8 [7, 8], creating 221,626 elements (Fig. 3). This type of element produced good results convergence, with acceptable computing times. It was during the meshing stage that the x, y, z baseline was set up; it will be retained throughout the experi- ments so that results can be read clearly. Mechanical characteristics of different elements All the rebuilt structures were considered individually as isotropes, that’s why Young’s modulus and Poisson’s coefficient of each structure were taken into account and incorporated in the model. The values applied to the different elements in the model are given in Table 1. Buchler et al. [6] showed that mechanical charac- teristics applied to muscle bodies and tendons had no effect on the model. The most important factor is still the surface and site of tendon insertion. Therefore, in Fig. 1 External contours of the proximal third of the humerus obtain with SURFDRIVER Fig. 2 The muscle bodies of the subscapular, supraspinatus are outlined and reconstructed 582 Surg Radiol Anat (2006) 28:581–587 123 this model we did not apply any specific Young’ modulus and Poisson’s coefficient to the structures. Once the model was discretized and its mechanical characteristics defined, the solver (MSC NASTRAN v7.5 MCS Software Corporation, Santa Ana, CA, USA) performed three levels of calculation to display the results: 1. To determine the matrices of rigidity for each element, so as to assemble them to obtain the global matrix of rigidity for the model and determine the vec- tor forces resulting from analysis of the load cases used. 2. To calculate the system solution in terms of dis- placement. 3. To calculate the other data such as stresses and deformation of the model. Loading conditions We loaded the humeral tubercles by simulating static abduction movement (considering only a gleno-hum- eral movement) in the plane of the scapula on our model. Two positions were studied: abduction at 10� (anatomical reference position) and abduction at 90� (maximum amplitude involving only the scapulohu- meral joint) (Fig. 4). Action was applied to each muscle in one direction only [25], according to the humeral position selected (Fig. 4). Points at the proximal and distal insertions for each muscle were defined using contours obtained via SURFDRIVER�. From these points, lines of action were created using PATRAN�. The muscular forces were applied along these lines. The load was applied at a single node (Fig. 4), then at multiple nodes (Fig. 5), as it is anatomically, to avoid concentrating stresses at the nodes in the model [18]. However, the number of nodes was chosen to retain adequate amplitude for observing the results. Because this is a static model, a glue type interaction between the glenoid and the humerus was chosen for these two positions. Muscular forces applied to each Fig. 3 Meshing of the reconstructed humerus with HEX8 elements Table 1 Young’s modulus and Poisson’s coefficient applied to the different elements in the model Cortical bone [2] Cancellous bone [10, 12, 19] Sub-chondral bone [5, 14, 21] Articular cartilage [2] ‘‘Young’’ bone ‘‘Old’’ bone ‘‘Young’’ bone ‘‘Old’’ bone ‘‘Young’’ bone ‘‘Old’’ bone ‘‘Young’’ bone ‘‘Old’’ bone Young’s modulus 16 GPa 12 GPa 550 MPa 250 MPa 3.5 GPa 3.5 GPa 2 MPa 2 MPa Poisson’scoefficient 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 Young’s modulus are obtained from specific studies of cortical and cancellous bone of the upper end of the humerus Surg Radiol Anat (2006) 28:581–587 583 123 muscle for an abduction position of 10 and 90� are given in Table 2 [26]. As our model refers to static arm positions, only the forces developed in these positions are taken into account [26]. Displaying the results given by the simulations To display the results, the maximum values of von Mises equivalent stresses (MPa) and total maximum deformation force (J mm–3) in cortical and cancellous bone in the models tested, were taken into account each in turn. We displayed the total densities of deformation forces so as to compare them. They were then extrapolated to deduce the resulting deformations. Indeed, for an element of volume V, subject to a range of stresses (rx, ry, rz, sxy, syz, szx) and a deformation field (�x, �y, �z, cxy, cyz, czx), the deformation force is: U ¼ 1 2 R V ðrxex þ ryey þ rzez þ sxycxy þ syzcyz þ szxczxÞ dV:The total density of deformation forces is then obtained by dividing this force U by the volume of the element. Results Abduction positions The type of force applied in the multinode model did not give a usable value for stress in cortical bone, which is why only the single node model was used. In this configuration, the stress noted for 90� abduction in ‘‘young’’ cortical bone was 0.72 MPa, whereas it was negligible at 10� abduction (Fig. 6). A small part of the transferred stress was recorded in the cancellous bone. The comparison of values according to the positions studied demonstrates that the reference position (arm the length of the body at 10� abduction in the plane of the scapula) exerts almost no force on the humeral tubercles. Fig. 4 The humeral tubercles are loaded by simulating static abduction movement in the plane of the scapula: abduction at 10� (anatomical reference position) and abduction at 90� (maximum amplitude involving only the scapulohumeral joint) Fig. 5 The muscular forces were applied at multiple nodes of the tubercles Table 2 The muscular forces applied to each muscle for an abduction position of 10� and 90� [26] Deltoideus Subscapularis Supraspinatus Infraspinatus/teres minor 10� of abduction 4.5 N 2.5 N 1 N 2 N 90� of abduction 116 N 70 N 25 N 60 N 584 Surg Radiol Anat (2006) 28:581–587 123 Comparison of the distribution of stresses according to type of bone The study of distribution of the total densities of deformation forces according to the type of bone mod- eled, shows that these values are higher for ‘‘old’’ bone than for ‘‘young’’ bone, which implies that there is more deformation in an osteoporotic bone under constant force (Fig. 7). The maximum stress values were re- corded in cortical bone. The stress values recorded varied considerably between the two positions studied, with a maximum load at 90� abduction (Fig. 8). Looking now at the depth of penetration of stresses in cancellous bone (sections at the subscapularis mus- cle insertion level), two observations can be made. Firstly, it can be seen that in the multinode model, the depth of penetration is fairly shallow (4 mm maxi- mum), and distributed over a wide area (Fig. 9). Sec- ondly, the depth of stress penetration is similar for both types of bone. Discussion The finite elements method involves evaluating dis- placements for all points of the structure being studied, by resolution of linear systems. From these displace- ments, deformations are obtained by derivation. The stresses are then evaluated from the deformations using a matrix characteristic of the structural material. Fig. 6 The type of force applied in the multinode model (left) did not give a usable value for stress in cortical bone, which is why only the single node model was used (right) Fig. 7 Comparison of the total densities of deformation forces (10–4 J mm–3) according to the type of bone modeled (‘‘old’’ bone and ‘‘young’’ bone) Fig. 8 The maximum stress values were recorded in cortical bone at 90� abduction Fig. 9 Depth of penetration of stresses in cancellous bone; the depth of penetration is fairly shallow, and distributed over a wide area Surg Radiol Anat (2006) 28:581–587 585 123 The aim of this study was to develop a digital model of the humerus, taking into account all the elements constituting the bone (cortical and cancel- lous bone, subchondral bone and articular cartilage), so as to perform a large number of non-destructive tests on anatomical specimens and prepare more accurate and fewer in-vitro tests. The acquisition of accurate geometry for the tendons and the bones is a fundamental requirement for the construction of three-dimensional FE models of joints. CT-scan yield superior spatial resolution than MRI and by this way provides excellent images of bones. Matching CT-scan views and anatomic preparations is of interest to differentiate between each tendon and surrounding bone and soft-tissue structures. In this study, a model of the proximal end of the humerus was developed to study the distribution of stresses in the humeral tubercles for two arm positions, using the forces representing muscle action in the rotator cuff muscles. We also studied the effect of bone quality on these data. This knowledge is important in so far as, by recreating an accurate anatomy of muscle bodies and their humeral insertions, we were able to describe exactly the mode and areas of diffusion of stresses in cortical and cancellous bone. This structural distinction is important, because the laws of stress distribution depend on regional variations in the moduli of elasticity. The von Mises stress distribution observed in this model was correlated with clinical observations. The regions in which a stress peak is observed correspond clinically to areas where tubercles are torn and frac- tured. The stresses noted in bone material did not ex- ceed 6 MPa, which matches results already published based on in-vivo studies [11, 17], or evaluated using digital models [2, 8]. In osteoporotic bone, the load on cortical bone is increased, as is the area of distribution of stress, an idea already put forward by Maldonado et al. [18]. Knowledge of the load on the bone is important in understanding the physiopathology of fractures and treating them. One of the aims of this study was to demonstrate the effect of bone quality on bone load, using a reference bone relative to an osteoporotic bone. The increase in load and its distribution for an osteoporotic bone must be taken clearly into consid- eration in designing osteosynthetic implants for use in treating fractures of the proximal end of the humerus, in which cancellous screws increase the stress on defi- cient cancellous tissue, by reducing the stresses nor- mally absorbed by the cortical bone. Similarly, it is important to consider bone quality when planning humeral reconstruction surgery. Criticisms of the model As for all numerical models, the boundary conditions are a difficult task. In the case of the shoulder, it is virtually impossible to reproduce exactly the real boundary conditions with all the muscles acting together. For these reason, the boundary condition was chosen to reproduce pure abduction. This is also why we only considered the rotator cuff muscles. However, the muscles modeled are the most important ones involved in abduction, which makes the investigation possible, particularly when using a multinode appli- cation, because of the complexity of calculations. Because of this, modeling mechanical actions using this digital model is fairly simplistic: indeed, modeling the line of action of muscles using a rope stretched between two points is still approximate. However, it helps validate the procedure used, producing the first results so as to envisage gradual improvement. The von Mises criterion could also beused in this study, because we considered our materials to be isotropic, although in reality they are not [1, 9, 12]. Conclusion The simulations performed in this study led to the development of an anatomically accurate digital model of the humerus, used to understand, display and map the distribution of stresses in the major and minor humeral tubercles generated by muscle tendons in the rotator cuff muscles during abduction. The results obtained were used to validate the model, by comparing them with published data, and to envisage more complex simulations. We can now incorporate prosthetic parts into the model to determine the maximum forces the material can withstand without risk of rupture, and complete the model with the scapula to evaluate the distribution and amplitude of stresses in the joint contact zone. Finally, this study demonstrates that, under physiological load on the rotator cuff, maximum stresses are observed at 90� abduction. The amplitude and dis- tribution of stresses is influenced mainly by bone quality. References 1. Anglin C, Tolhurst P, Wyss U, Pichora D (1999) Glenoid cancellous bone strength and modulus. J Biomech 32:1091– 1097 2. Barea C (1998) Modelisation 3D par la méthode des élé- ments finis d’une articulation scapulo-humérale. Application à l’étude des contraintes sur une épaule saine et avec diffé- rents types de glènes prothétiques. Thesis. Université Paul Sabatier, Toulouse 586 Surg Radiol Anat (2006) 28:581–587 123 3. Barea C, Hobatho M, Darmana R, Mansat P, Mansat M (1998) Three dimensional finite element modelling of normal and implanted glenoid: validations and simulations. In: 11th Conference of the ESB. Toulouse, p 27 4. Biluart F, Gagey O, Skalli W, Mitton D (2006) Biomechanics of the deltoideus. Surg Radiol Anat 28:76–81 5. Braidotti P, Bemporade E, D’Alessiot T, Sciuto S, Stagni I (2000) Tensile experiments and SEM fractography on bovine subchondral bone. J Biomech 9:1153–1157 6. Büchler P, Ramaniraka N, Rakotomanana L, Iannotti J, Farron A (2002) A finite element model of the shoulder: application to the comparison of normal and osteoarthritic joints. Clin Biomech 17:630–639 7. Büchler P, Ramaniraka N, Rakotomanana L, Iannotti J, Farron A (2002) A finite model of the shoulder: application to the comparison of normal and osteoarthritic joints. Clin Biomech 17:630–639 8. Couteau B, Mansat P, Estivalezes E, Darmana R, Mansat M, Egan J (2001) Finite element analysis of the mechanical behavior of a scapula implanted with a glenoid prosthesis. Clin Biomech (Bristol, Avon) 16:566–575 9. Frich L (1997) Bone strength and materials properties of the glenoid. J Shoulder Elbow Surg 6:97–104 10. Friedman R (1992) Finite element of the glenoid component: effect of the design parameters on stress distribution. J Shoulder Elbow Surg 1:261–270 11. Frost H (1987) The mechanostat: a proposed pathogenic mechanism of osteoporoses and the bone mass effects of mechanical and non-mechanical agents. Bone Minerology 2:73–85 12. Hobatho M, Rho J, Ashman R (1992) Atlas of mechanical properties of human cortical and cancellous bone. In vivo assessment of bone quality by vibration and wave propaga- tion techniques. Part II. Van der Perre G, Lowet G, Borg- wardt A (eds) ACCO, Leuven 13. Hughes R, An K (1996) Force analysis of rotator cuff mus- cles. Clin Orthop 330:75–83 14. Huiskes R (1987) Finite element analysis of acetabular reconstruction. Non cemented threated cups. Acta Orthop Scand 58:620–625 15. Karlsson D, Peterson B (1992) Towards a model for force predictions in the human shoulder. J Biomech 25:189–199 16. Lacroix D, Prendergast P (1997) Stress analysis of glenoid component designs for shoulder arthroplasty. Proc Inst Mech Eng 211:467–474 17. Lanyon L, Hampson W, Goodship A, Shan J (1975) Bone deformation recorded in vivo from strain gauges attached to the human tibial shaft. Acta Orthop Scand 46:256–268 18. Maldonado Z, Seebeck J, Heller M, Brandt D, Hepp P, Lill H, Duda G (2003) Straining of the intact and fractured proximal humerus under physiological-like loading. J Bio- mech 36:1865–1873 19. Mansat P (1996) Propriétés mécaniques osseuse de la glène. Application à la modélisation tridimensionnelle par la méthode des éléments finis. Thesis. Université Paul Sabatier, Toulouse 20. Orr T, Carter D, Schurman D (1988) Stress analyses of gle- noid component designs. Clin Orthop 232:217–224 21. Pedersen D, Crowninshield R, Brand R, Johnston R (1982) An axisymetric model of acetabular components in total hip arthroplasty. J Biomech 15:305–315 22. Stone K, Grabowski J, Cofield R, Morrey B, An K (1999) Stress analyses of glenoid components in total shoulder arthroplasty. J Shoulder Elbow Surg 8:151–158 23. Van der Helm F (1994) Analysis of the kinematic and dynamic behavior of the shoulder mechanism. J Biomech 27:527–550 24. Van der Helm F (1994) A finite element musculoskeletal model of the shoulder mechanism. J Biomech 27:551–569 25. Van der Helm F, Veenbaas R (1991) Modelling the mechanical effect of muscles with large attachment sites: application to the shoulder mechanism. J Biomech 21:1151– 1163 26. Wuelker N, Wirth CJ, Plitz W, Roetman B (1995) A dynamic shoulder model: reliability testing and muscle force study. J Biomech 28:489–499 Surg Radiol Anat (2006) 28:581–587 587 123 Finite element analysis of the strain distribution in the humeral head tubercles during abduction: comparison of young �and osteoporotic bone Abstract Introduction Materials and methods Obtaining anatomical shapes in 3D Meshing Mechanical characteristics of different elements Fig1 Fig2 Loading conditions Fig3 Tab1 Displaying the results given by the simulations Results Abduction positions Fig4 Fig5 Tab2 Comparison of the distribution of stresses according to type of bone Discussion Fig6 Fig7 Fig8 Fig9 Criticisms of the model Conclusion References > /ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 150 /GrayImageDepth -1 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true/GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 600 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile (None) /PDFXOutputCondition () /PDFXRegistryName (http://www.color.org?) /PDFXTrapped /False /Description /ENU >> >> setdistillerparams > setpagedevice